Abstract
Keywords
Magnetized plasma and magnetic flux are widely researched in controlled nuclear fusion[
1. Relativistic speed electron beams with ultrashort bunch length will penetrate specimen with millimeter size in several tens of picoseconds, which ensures the quasi-static state of the diagnosed specimen and makes it possible to observe ultrafast phenomena.
2. Compared with the other relativistic charged particles, electron has lower magnetic rigidity, which makes it more sensitive to the electromagnetic field.
3. Besides, ultrashort relativistic electron bunches can be generated and manipulated easier than other charged particles like proton or ion.
High energy electron beams as the probe of the high energy density matter (HEDM) diagnosis was proposed in the past several years[
1 Principle of dual degrees of freedom diagnosis
1.1 Electron-target interaction
When penetrating through the diagnosed system, the electrons are mainly affected by two kinds of interactions: multiple Coulomb scattering and E/B field deflection. We think that the interactions of electrons with matter and electromagnetic field are independent, the matter makes electrons scatter randomly and the field makes electrons deflect to a specific direction.
Electron beams from accelerator is very close to parallel monoenergetic beams. The distribution of multiple Coulomb scattering angles can be described by the following function:
where
Here φ is the scattering angle of electrons, p is the momentum,
The deflection angle from E/B field can be described by the following equation:
where
1.2 Beam optics theory
In beam optics, the transportation of monoenergetic electron beams can be expressed by the following matrix equation:
where
The Fourier planes are where the transport matrix parameters R11F or R33F is 0, the footnote F indicates that the matrix is from the object plane to the Fourier plane. If the Fourier planes of x-axis and y-axis are coincident along the z-axis by optimization, the following matrix equation can be satisfied:
It means that the electron angle information at the object plane has been translated into the position information at this plane. An aperture at this plane can achieve angle selection for passing electrons. The principle schematic is shown in Fig.1.
Figure 1.Schematic diagram of HEELR to make angle selection with an aperture. (a) The schematic of point-to-point imaging beamline with an aperture at the Fourier plane. (b) The angle distribution of incident electrons. (c1) and (c2) illustrate the electron angle distributions after target with different thicknesses and the corresponding position distributions at the Fourier plane. (d1) and (d2) are the electron angle distributions after penetrating the system with E/B field and the corresponding position distributions at the Fourier plane. The purple rectangular shadow indicates the aperture acceptance area
In the x-axis direction, the spatial resolution is
1.3 Introduction of areal density difference method
In view of the above analysis, the transmittance of monoenergetic electron beams is dominated by three factors, which are the matter areal density, the E/B field strength and the aperture size. For the multiple degrees of freedom diagnosis, more equations are necessary. In single shot experiment, the aperture size is unchangeable, and
Figure 2.The principle of areal density difference method to diagnose the areal density and E/B field. (a) The scattering angle distributions after the grille scattering target. Different color corresponds to different thicknesses (
With the grille scattering target, two transmittance values can be obtained for each image point in single radiography. One is the measured one, the other is got from fitting the neighbor area, which has different areal density at the scattering target. The corresponding transmittances at the image plane can be denoted by Tr(t+t1,
Extend to two-dimensional plane, the transmittance density can be described as[
where
where
where
When the aperture is set as a ring, the inner and outer radii satisfy
The transmittance will still be independent of the direction of
In two-dimensional plane, we can set a pre-modulation scattering target with four thicknesses before the real diagnosed target, as shown in Fig.3(a). Fig.3(b) is the front view of a scattering target example.
Figure 3.Overall design of the dual degrees of freedom diagnostic. (a) The incident beams will be scattered by the scattering target before penetrating the real diagnosed target. (b) The front view (perpendicular to the beam bunches) of the scattering target, different color indicates different thickness
Thus, four transmittance values can be obtained for each point at the image plane. The first one is the measured one, the other three transmittances is derived from fitting and interpolating the transmittance along the three lines in Fig.3(b). Finally, four equations can be obtained as Eq.(10), which is more than enough to give a precise solution of t and
2 Simulation design and results analysis
To analyze the feasibility of this method, we simulated 50 MeV electrons diagnosing hydrogen and E/B field. To make the areal density and E/B field have independence in separate dimensions, we set the diagnosed sample as Fig.4(a).
Figure 4.(a) The wedge shape represents the hydrogen sample with gradient areal density. The color indicates the value of deflection angle from the E/B field and the coordinate system is as shown. (b) The ellipse shape aperture sets the upper limit of the angle acceptance as 1mrad. (c) The ring shape aperture which sets the angle acceptance as (1,2) mrad
The sample was set as a wedge shape. Considering the diagnosis ability of 50 MeV electron beams, we set the areal density of hydrogen increasing from 0.5 mg/cm2 to 5.5 mg/cm2 along the x-axis and the density was 1 g/cm3, θ was decreasing from 5 mrad to 0 mrad along the y-axis. We tried a set of scattering target thicknesses, 0 µm, 2 µm, 4 µm and 8 µm, for contrast, so there were
Figure 5.The analysis results when the angle acceptance of the ellipse shape aperture is 1 mrad
As shown in Fig.5(a), for the transmittance of a point at the image plane, there will be a corresponding contour line with X-θ as variables. Draw the contour lines with different scattering target thicknesses in one picture, as illustrated in Fig.5(b), there will be an intersection point between the two contour lines with the same transmittance, which indicates the solution (X, θ). The resolution of X and θ is better in larger θ area. In lower θ area, the contour lines are almost parallel. This indicates that this design is suitable for diagnosing the fluid in strong E/B field. Meanwhile, compare different combinations in Fig.5(b), it can be concluded that the larger the difference of scattering target thicknesses is, the better resolution of (X,θ) we can get. In other words, contour lines of 0 µm and 8 µm are the better choice to fix the solution (X, θ). The other two thicknesses can improve the E/B field diagnosis range. When the aperture is set as an ellipse hole, as mentioned above, the resolution will deteriorate in low θ area. To solve this problem, a ring aperture as Fig.4(c) demonstrates, which sets the angle acceptance as (1,2) mrad, is adopted. The analysis results are shown in Fig.6.
In Fig.6(a), when the angle acceptance range is set to (1, 2) mrad, the transmittance distributions have obviously been changed. Compare Fig.6(b) with Fig.5(b), the resolution at low θ area is improved a lot. On the other hand, two contour lines with the same transmittance in Fig.6(b) could have two intersection points, which correspond to two different solutions (X01, θ01) and (X02, θ02). Under such circumstances, a third contour line from different scattering target thickness but with the same transmittance can help to make the solution settled.
Figure 6.The analysis results when the angle acceptance is set as from 1mrad to 2 mrad by a ring shape aperture
3 Conclusion
In summary, the areal density difference diagnostic method based on HEELR is proposed to make dual degrees of freedom consists of areal density and E/B field strength diagnosis in this paper. In two-dimensional plane, four transmittances from different scattering target thicknesses can be obtained. When the angle acceptance area of the beam line is a circle, this method is suitable for the diagnosed system with strong E/B field. When the aperture is set as a ring, the effective diagnostic range can be extended to the low E/B field area. In an actual experiment, we need to adjust the scattering target thickness, the aperture shape and size, even the incident beam energy to match different diagnosed system. Furthermore, since this method can provide four equations, the vector of field and another degree of freedom may be obtained, just like the coupling in the strong coupled system diagnosing.
References
[1] Walsh C A, Chittenden J P, McGlinchey K, et al. Self-generated magnetic fields in the stagnation phase of indirect-drive implosions on the National Ignition Facility[J]. Physical Review Letters, 118, 155001(2017).
[2] Fox W, Bhattacharjee A, Germaschewski K. Magnetic reconnection in high-energy-density laser-produced plasmas[J]. Physics of Plasmas, 19, 056309(2012).
[3] Gotchev O V, Chang P Y, Knauer J P, et al. Laser-driven magnetic-flux compression in high-energy-density plasmas[J]. Physical Review Letters, 103, 215004(2009).
[4] Lindemuth L R, Ekdahl C A, Fowler C M, et al. US/Russian collaboration in high-energy-density physics using high-explosive pulsed power: ultrahigh current experiments, ultrahigh magnetic field applications, and progress toward controlled thermonuclear fusion[J]. IEEE Transactions on Plasma Science, 25, 1357-1372(1997).
[5] Merrill F E, Campos E, Espinoza C, et al. Magnifying lens for 800 MeV proton radiography[J]. Review of Scientific Instruments, 82, 103709(2011).
[6] Kantsyrev A V, Golubev A A, Bogdanov A V, et al. TWAC-ITEP proton microscopy facility[J]. Instruments and Experimental Techniques, 57, 1-10(2014).
[7] Rygg J R, Séguin F H, Li C K, et al. Proton radiography of inertial fusion implosions[J]. Science, 319, 1223-1225(2008).
[8] Tommasini R, Landen O L, Hopkins L B, et al. Time-resolved fuel density profiles of the stagnation phase of indirect-drive inertial confinement implosions[J]. Physical Review Letters, 125, 155003(2020).
[9] Martynenko A S, Pikuz S A, Skobelev I Y, et al. Optimization of a laser plasma-based X-ray source according to WDM absorption spectroscopy requirements[J]. Matter and Radiation at Extremes, 6, 014405(2021).
[10] Zhao Yongtao, Zhang Zimin, Gai Wei, et al. High energy electron radiography scheme with high spatial and temporal resolution in three dimension based on a e-LINAC[J]. Laser and Particle Beams, 34, 338-342(2016).
[11] Xiao Jiahao, Zhang Zimin, Cao Shuchun, et al. Areal density and spatial resolution of high energy electron radiography[J]. Chinese Physics B, 27, 035202(2018).
[12] Wang Feng, Jiang Shaoen, Ding Yongkun, et al. Recent diagnostic developments at the 100 kJ-level laser facility in China[J]. Matter and Radiation at Extremes, 5, 035201(2020).
[13] Li C K, Séguin F H, Frenje J A, et al. Measuring
[14] Li C K, Séguin F H, Frenje J A, et al. Monoenergetic-proton-radiography measurements of implosion dynamics in direct-drive inertial-confinement fusion[J]. Physical Review Letters, 100, 225001(2008).
[15] Liao Guoqian, Li Yutong, Zhu Baojun, et al. Proton radiography of magnetic fields generated with an open-ended coil driven by high power laser pulses[J]. Matter and Radiation at Extremes, 1, 187-191(2016).
[16] Schumaker W, Nakanii N, McGuffey C, et al. Ultrafast electron radiography of magnetic fields in high-intensity laser-solid interactions[J]. Physical Review Letters, 110, 015003(2013).
[17] Zhu P F, Zhang Z C, Chen L, et al. Ultrashort electron pulses as a four-dimensional diagnosis of plasma dynamics[J]. Review of Scientific Instruments, 81, 103505(2010).
[18] Li Junjie, Wang Xuan, Chen Zhaoyang, et al. Ultrafast electron beam imaging of femtosecond laser-induced plasma dynamics[J]. Journal of Applied Physics, 107, 083305(2010).
[19] Chen Long, Li Runze, Chen Jie, et al. Mapping transient electric fields with picosecond electron bunches[J]. Proceedings of the National Academy of Sciences of the United States of America, 112, 14479-14483(2015).
[20] Merrill F, Harmon F, Hunt A, et al. Electron radiography[J]. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 261, 382-386(2007).
[21] Merrill F E, Goett J, Gibbs J W, et al. Demonstration of transmission high energy electron microscopy[J]. Applied Physics Letters, 112, 144103(2018).
[22] Zhou Zheng, Fang Yu, Chen Han, et al. Visualizing the melting processes in ultrashort intense laser triggered gold mesh with high energy electron radiography[J]. Matter and Radiation at Extremes, 4, 065402(2019).
[23] Xiao Jiahao, Du Yingchao, Zhang Shizheng, et al. Ultrafast high-energy electron radiography application in magnetic field delicate structure measurement[J]. Laser and Particle Beams, 6683245(2021).
[24] Xiao Jiahao, Du Yingchao, Li Haoqing, et al. Ultrafast high energy electron lens radiography suitable f transient electromagic field diagnosis [J]. Journal of Instrumentation, 2022, 17(01):P01033 DOI:10.1088174802211701P01033
[25] Hirayama H, Namito Y, Bielajew A F, et al. The EGS5 code system[R]. SLACR730, 2005.
Set citation alerts for the article
Please enter your email address