• Acta Optica Sinica
  • Vol. 42, Issue 14, 1412002 (2022)
Cong Sun1、2、*, Mengna Jia3, and Qifeng Yu4
Author Affiliations
  • 1State Key Laboratory of Astronautic Dynamics, Xi′an710043, Shaanxi , China
  • 2Xi′an Satellite Control Center, Xi′an710043, Shaanxi , China
  • 3China Academy of Space Technology (Xi′an), Xi′an710000, Shaanxi , China
  • 4College of Aeronautics and Astronautics, National University of Defense Technology, Changsha 410073, Hunan , China
  • show less
    DOI: 10.3788/AOS202242.1412002 Cite this Article Set citation alerts
    Cong Sun, Mengna Jia, Qifeng Yu. Calibration and Epipolar Rectification Methods for Stereo Scheimpflug Cameras[J]. Acta Optica Sinica, 2022, 42(14): 1412002 Copy Citation Text show less

    Abstract

    Aiming at the stereo imaging system with Scheimpflug cameras, the calibration and epipolar rectification methods are investigated in this paper. The epipolar geometry of stereo Scheimpflug cameras is described, on the basis of which a robust and effective stepwise calibration method is proposed. Based on the calibration results of intrinsic and extrinsic parameters, the epipolar rectification model of stereo Scheimpflug cameras is presented. Furthermore, a three-dimensional digital image correlation (DIC) deformation measurement system is developed based on Scheimpflug cameras. The experimental results show that the accuracy of the three-dimensional DIC deformation measurement system is superior to 10 μm, which demonstrates the effectiveness and accuracy of the stepwise calibration and epipolar rectification of proposed stereo Scheimpflug cameras.
    Cong Sun, Mengna Jia, Qifeng Yu. Calibration and Epipolar Rectification Methods for Stereo Scheimpflug Cameras[J]. Acta Optica Sinica, 2022, 42(14): 1412002
    Download Citation