• Acta Optica Sinica
  • Vol. 39, Issue 3, 0312002 (2019)
Rui Zhang1、2、3、*, Yuanyuan Chen1、2、3, Ning Jing1、2、3, Zhibin Wang1、2、3, Kewu Li1、2、3, and Kunyang Xie1、2、3
Author Affiliations
  • 1 Engineering and Technology Research Center of Shanxi Province for Optical-Electric Information and Instrument, North University of China, Taiyuan, Shanxi 0 30051, China
  • 2 Key Laboratory of Instrument Science & Dynamic Measurement, Ministry of Education, North University of China, Taiyuan, Shanxi 0 30051, China;
  • 3 State Key Laboratory for Electronic Measurement Technology, North University of China, Taiyuan, Shanxi 0 30051, China
  • show less
    DOI: 10.3788/AOS201939.0312002 Cite this Article Set citation alerts
    Rui Zhang, Yuanyuan Chen, Ning Jing, Zhibin Wang, Kewu Li, Kunyang Xie. High-Precision Measurement of Mid-Infrared Waveplate Phase Retardation Based on Dual Photoelastic Difference Frequency Modulation[J]. Acta Optica Sinica, 2019, 39(3): 0312002 Copy Citation Text show less
    Schematic of mid-infrared waveplate retardation measurement based on dual-PEMs difference frequency modulation
    Fig. 1. Schematic of mid-infrared waveplate retardation measurement based on dual-PEMs difference frequency modulation
    Schematic of mid-infrared waveplate retardation imaging based on dual-PEMs difference frequency modulation
    Fig. 2. Schematic of mid-infrared waveplate retardation imaging based on dual-PEMs difference frequency modulation
    PEM size and simulation results. (a) PEM size; (b) vibration mode of PEM; (c) peak retardation amplitude distribution of PEM
    Fig. 3. PEM size and simulation results. (a) PEM size; (b) vibration mode of PEM; (c) peak retardation amplitude distribution of PEM
    Experimental setup
    Fig. 4. Experimental setup
    Experiment results obtained by the NI card. (a) Without waveplate; (b) with waveplate
    Fig. 5. Experiment results obtained by the NI card. (a) Without waveplate; (b) with waveplate
    Experiment results of system sensitivity. (a) LCVR retardation values at different driving voltages; (b) experiment results of the system at different LCVR driving voltages
    Fig. 6. Experiment results of system sensitivity. (a) LCVR retardation values at different driving voltages; (b) experiment results of the system at different LCVR driving voltages
    Phase retardation amplitude of dual-PEMsδ01PEM1, δ01PEM2δ02PEM1, δ02PEM2δ03PEM1, δ03PEM2δ04PEM1, δ04PEM2
    Experiment result δX/rad1.571151.571051.571161.5709
    Ellipsometer measurement result δX/rad1.5711
    Relative error /%0.0030.0030.0040.0006
    Table 1. Comparison of experimental results and waveplate detector measurement results
    Rui Zhang, Yuanyuan Chen, Ning Jing, Zhibin Wang, Kewu Li, Kunyang Xie. High-Precision Measurement of Mid-Infrared Waveplate Phase Retardation Based on Dual Photoelastic Difference Frequency Modulation[J]. Acta Optica Sinica, 2019, 39(3): 0312002
    Download Citation