• Spectroscopy and Spectral Analysis
  • Vol. 42, Issue 9, 2870 (2022)
Fei ZHANG1、1; 5;, Xia HUA2、2; *;, Jia-ying YUAN3、3;, Fan YOU1、1;, Ren-cai YE4、4;, Li DING4、4;, and Jian-mei ZHAO4、4;
Author Affiliations
  • 11. Sichuan Center for Disease Control and Prevention, Chengdu 610041, China
  • 22. Sichuan Center for Food and Drug Evaluation, Inspection & Monitoring, Chengdu 610017, China;
  • 33. Chengdu University of Traditional Chinese Medicine, Chengdu 610032, China
  • 44. Liangshan Zhou Center for Disease Control and Prevention, Liangshan 615000, China
  • show less
    DOI: 10.3964/j.issn.1000-0593(2022)09-2870-05 Cite this Article
    Fei ZHANG, Xia HUA, Jia-ying YUAN, Fan YOU, Ren-cai YE, Li DING, Jian-mei ZHAO. Determination of Thallium in Blood of Occupational Exposed Population by Inductively Coupled Plasma Mass Spectrometry With High Matrix Introduction[J]. Spectroscopy and Spectral Analysis, 2022, 42(9): 2870 Copy Citation Text show less
    Mass spectrogram of standard series and internal standard (a); Standard curve equation of thallium (b)
    Fig. 1. Mass spectrogram of standard series and internal standard (a); Standard curve equation of thallium (b)
    Variation of internal standard element signals
    Fig. 2. Variation of internal standard element signals
    参数设定值
    RF power1 550 W
    Cool gas flow15.0 L·min-1
    Auxiliary gas flow0.90 L·min-1
    Plasma gas flow15 L·min-1
    Nebulizer pump speed0.1 r·min-1
    Carrier gas flow rate1.05 L·min-1
    Acquisition modeMass spectrogram
    Tuning modeNo Gas
    Plasma modeHMI
    Table 1. Instrumental parameters for HMI-ICP-MS
    冻干全血质
    量控制样品
    方法测定值/(μg·L-1)精密度/%误差/%参考值范围/(μg·L-1)
    111.7, 10.9, 11.0, 11.4, 11.0, 12.86.2811.9
    211.0, 10.5, 11.5, 10.2, 10.7, 9.755.774.79
    L-2310.6, 10.8, 10.2, 9.91, 10.7, 10.63.263.5210.1(8.1~12.1)
    411.2, 11.2, 11.9, 10.9, 11.6, 10.83.7110.4
    510.3, 11.1, 11.3, 11.1, 10.7, 10.24.256.34
    132.7, 31.5, 34.4, 33.8, 30.9, 34.24.453.39
    234.4, 33.9, 31.1, 33.8, 34.7, 32.64.024.84
    L-3333.4, 31.9, 33.8, 31.8, 32.4, 32.52.472.5531.8(29.0~34.5)
    433.8, 32.9, 34.3, 34.5, 31.3, 32.23.804.12
    533.5, 34.5, 32.1, 33.1, 34.2, 33.22.564.89
    Table 2. Analytical results of thallium in blood under different treatments by HMI-ICP-MS (n=6)
    本底
    /(μg·L-1)
    加标
    /(μg·L-1)
    测得值/(μg·L-1)回收率
    /%
    批内精密度
    /%
    批间精密度
    /%
    <0.422.502.63, 2.70, 2.60, 2.59, 2.55, 2.61, 2.49103.82.52-
    2.71, 2.45, 2.61, 2.44, 2.57, 2.38, 2.64101.7-4.77
    <0.4215.015.2, 15.7, 16.3, 15.3, 15.2, 15.6, 16.1104.22.81-
    15.4, 14.0, 14.2, 15.4, 14.9, 15.1, 13.992.7-4.44
    <0.4275.075.6, 74.8, 73.7, 72.9, 75.2, 72.1, 74.398.81.71-
    75.4, 76.5, 77.3, 72.9, 71.2, 74.6, 73.699.3-2.84
    Table 3. Recovery and precision of the method (n=7)
    Fei ZHANG, Xia HUA, Jia-ying YUAN, Fan YOU, Ren-cai YE, Li DING, Jian-mei ZHAO. Determination of Thallium in Blood of Occupational Exposed Population by Inductively Coupled Plasma Mass Spectrometry With High Matrix Introduction[J]. Spectroscopy and Spectral Analysis, 2022, 42(9): 2870
    Download Citation