• Laser & Optoelectronics Progress
  • Vol. 58, Issue 3, 3140051 (2021)
Wang Qiubu1, Liu Wenqing1、2, Zhang Tianshu1、2、*, and Fu Yibin2
Author Affiliations
  • 1Institute of Physical Science and Information Technology, Anhui University, Hefei , Anhui 230601, China
  • 2Key Laboratory of Environment Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei , Anhui 230031, China
  • show less
    DOI: 10.3788/LOP202158.0314005 Cite this Article Set citation alerts
    Wang Qiubu, Liu Wenqing, Zhang Tianshu, Fu Yibin. Design of High Sensitive Power Detection System for Pulsed Laser[J]. Laser & Optoelectronics Progress, 2021, 58(3): 3140051 Copy Citation Text show less
    System overall design block diagram
    Fig. 1. System overall design block diagram
    Photoelectric conversion circuit
    Fig. 2. Photoelectric conversion circuit
    Peak hold and discharge circuit
    Fig. 3. Peak hold and discharge circuit
    Synchronous trigger circuit
    Fig. 4. Synchronous trigger circuit
    Block diagram of program execution process
    Fig. 5. Block diagram of program execution process
    Timing diagram of acquisition process
    Fig. 6. Timing diagram of acquisition process
    Oscilloscope real-time monitoring graph
    Fig. 7. Oscilloscope real-time monitoring graph
    Comparison of measured peak output and ideal value
    Fig. 8. Comparison of measured peak output and ideal value
    Schematic of experimental device
    Fig. 9. Schematic of experimental device
    Linear fitting result of power and voltage
    Fig. 10. Linear fitting result of power and voltage
    SymbolValue
    Spectral response range /nm350-1100
    Dark current /nA0.25
    Responsivity /(A·W-1)0.35(λ=532 nm)
    Rise time /ns70
    Active area /(mm×mm)3.504×3.504
    Table 1. PS13-6b main parameters
    Part numberBand width /MHzTransconductance /(mA·V-1)Slew rate /(V·μs-1)Supply voltage /V
    OPA8608095900±5
    OPA615710722500±5
    OPA6608501253000±5
    MAX436200-850±5
    Table 2. Comparison of common transconductance amplifier parameters
    Designed power meter /WStandard power meter /WError rate /%
    0.3480.3334.5
    0.4740.4534.6
    0.6300.6054.1
    0.7810.7652.1
    0.9280.9121.8
    1.0261.0431.6
    1.1161.1351.7
    1.1741.2032.4
    Table 3. Comparison of errors between homemade power detection system and standard power meter
    Wang Qiubu, Liu Wenqing, Zhang Tianshu, Fu Yibin. Design of High Sensitive Power Detection System for Pulsed Laser[J]. Laser & Optoelectronics Progress, 2021, 58(3): 3140051
    Download Citation