• Acta Optica Sinica
  • Vol. 33, Issue 6, 612010 (2013)
Cao Shaoqian1、2、*, Bu Yang1, Wang Xiangzhao1, Li Sikun1, Tang Feilong1、2, and Li Zhongliang1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/aos201333.0612010 Cite this Article Set citation alerts
    Cao Shaoqian, Bu Yang, Wang Xiangzhao, Li Sikun, Tang Feilong, Li Zhongliang. Measurement Error Analysis for Mueller Matrix Based on a Single Photo-Elastic Modulator[J]. Acta Optica Sinica, 2013, 33(6): 612010 Copy Citation Text show less

    Abstract

    A measurement error equation and a relative error analysis method for the Mueller matrix measurement technique based on a single photo-elastic modulator are proposed. Considering the relative error and the condition number, two optimum angle sets for the quarter-wave plate are obtained, which diminish the maximum relative error of each element of the Mueller matrix. The experimental results show that the maximum relative errors of each element of the Mueller matrix of the measured quarter-wave plate are 0.12% and 0.20% respectively by using the two optimum angle sets above. Compared to the maximum relative error of 0.83% of the traditional optimum angle sets {-90°,-45°,30°,60°} for the quarter-wave plate, the maximum relative errors of each element of the Mueller matrix of the measured quarter-wave plate are decreased by 85.54% and 75.90% respectively by using the two optimum angle sets above.
    Cao Shaoqian, Bu Yang, Wang Xiangzhao, Li Sikun, Tang Feilong, Li Zhongliang. Measurement Error Analysis for Mueller Matrix Based on a Single Photo-Elastic Modulator[J]. Acta Optica Sinica, 2013, 33(6): 612010
    Download Citation