Epilepsy is a brain disorder that occurs due to hypersynchronous firing of neurons in the brain, an abnormal electrical activity of the brain. Epilepsy, characterized by involuntary body movements, loss of consciousness, and loss of bowel or bladder control, is the most common chronic neurological disorder[
Detecting epileptic seizures as early as possible, giving appropriate and reasonable treatment, and timely manual intervention before the seizure or recurrence, can bring more treatment possibilities to epilepsy patients and alleviate their symptoms. According to the abnormal repetitive neural discharge characteristics in the epileptic focus during the onset of epilepsy, doctors can detect the onset of epilepsy by observing the discharge phenomenon of the patient's brain. Electroencephalography (EEG) monitors the electrical function of the brain. Epilepsy can be effectively studied by generating multidimensional time series from EEG recordings that are complex, non-stationary, nonlinear, and random. EEG can be divided into diagnostic, localization, and monitoring applications. As an effective means of monitoring epileptic seizures, EEG is widely used clinically.
At present, such EEG analysis work is mainly through the doctor's visual inspection and manual labeling[
Generally speaking, patients with different types of illnesses have different symptoms during seizures. Therefore, it is extremely important to fully understand the onset patterns of epilepsy patients. However, the classification of epilepsy is a time-consuming and tedious task, which places a heavy burden on clinicians. In view of this, people are working hard to develop an automatic detection system to reduce the workload to assist neurologists in detecting EEG signals during epileptic seizures and relieve the burden on physicians.
A flowchart of EEG-based automatic epilepsy detection is shown in
Figure 1.(Color online) A flowchart of EEG-based automatic epilepsy detection.
EEG basics
EEG is a map of electrical potential activity that demonstrates spontaneous, continuous, and rhythmic electrical potential changes in neurons in the brain. It can display the physiological, psychological and disease information of the human body. Due to its low cost and real-time characteristics, EEG has been widely used in the medical field. So far, EEG technology has already become one of the important means of clinical detection.
The EEG signal is a non-stationary signal with the characteristics of suddenness and randomness. According to research, the analysis frequency range of EEG signals is between 0.5 and 100 Hz, and most of them are concentrated in 0.5−70 Hz. According to frequency, normal EEG signals can be divided into four bands: δ, θ, α, β and γ waves, where different frequency bands of EEG reflect different states of the brain.
(1) δ wave: It is mainly distributed in the frequency range between 0.5 and 3 Hz, and the amplitude of the signal is between 20 and 200 μV. Its amplitude is often the highest, while the waveform is the slowest. This waveform is mainly concentrated in the areas of the frontal and occipital lobes and occurs mainly during deep sleep or deep anesthesia.
(2) θ wave: It is mainly distributed in the frequency range between 4 and 7 Hz, and the amplitude of the signal is between 20 and 150 μV. According to research, this band mainly appears in the frontal and parietal regions and usually appears in early childhood. This waveform may occur during sleepiness and arousal phases in older children and adults. In addition, this wave can also appear when encountering difficulties or feeling depressed, which is primarily associated with relaxation, meditation and creativity.
(3) α wave: It is mainly distributed in the frequency range between 8 and 13 Hz, and the amplitude of the signal is between 20 and 100 μV. This waveform is different from the previous waveforms in the distribution range, and exists in multiple regions of the brain, but it is mainly concentrated in the top and occipital regions of the brain. This waveform appears when the eyes are closed or relaxed, and fades away when the eyes are open or mentally drained. When the state of thinking is suddenly interrupted by external stimuli, the wave will disappear quickly. This phenomenon is usually called the wave-blocking phenomenon.
(4) β wave: It is mainly distributed in the frequency range between 14 and 30 Hz, and the amplitude of the signal is between 5 and 20 μV. This wave mainly occurs in the frontal, parietal and central regions of the brain. This wave is often associated with active, busy or anxious thinking and active attention, and is strongly associated with motor behavior, attenuating during active movement. Studies have found that this wave often occurs during the day in a state of waking consciousness. At the same time, because it is related to imaginary EEG signals, it is of great significance in the field of EEG.
(5) γ wave: It is mainly distributed in the frequency band greater than 30 Hz, and the amplitude of the signal is less than 20 μV. This wave mainly occurs in the frontal and parietal regions of the brain, during extreme hyperactivity, or when neuronal excitation is greatly increased. However, it is generally considered that this wave has little to do with the characteristics of the EEG signal, so in the research of processing EEG signals, the high frequency is usually filtered.
Epilepsy is a chronic neurological disease caused by excessive discharge of central neurons. EEG signals during epileptic seizures are quite different from normal conditions. The waveform, amplitude, frequency and other characteristics of EEG signals are often used to distinguish epileptic signals from non-epileptic signals. At present, the most important way is to use the waveform changes in the EEG to detect epilepsy signals, so as to carry out further treatment. During the course of the research, it was found that sharp waves, spike waves, and multi-spike slow waves usually appear during epileptic seizures.
EEG acquisition equipment
As early as the late 18th century, scientists demonstrated that certain animals were capable of generating bioelectric waveforms[
The first human detection of EEG signals on the surface of the scalp was in 1924, when Hans Berger, Professor of Neurology at the University of Jena, Germany, discovered alpha brain waves from the scalp of young children. Initially, Berger used two platinum needle-like electrodes inserted into the cerebral cortex at the site of the subject's skull injury to acquire signals, but later, after repeated trials, it was demonstrated that EEG signals could be acquired through electrodes from the superficial skin of the scalp without the insertion of invasive electrodes, making this non-invasive EEG acquisition technique more suitable for wearable device applications.
In order to obtain the EEG signal, the sensor-electrode can be used to collect the signal and transmit the data to the back-end system for further processing[
The following summarizes the research status and development trend of the hardware system of EEG acquisition equipment, summarizes the basic structure of EEG acquisition equipment, and focuses on sorting out and discussing different methods for optimizing EEG acquisition equipment.
EEG acquisition equipment is a data source for various EEG applications and research. As a precision test instrument, it is necessary to ensure the safety of the user, as well as the high precision and high common mode rejection ratio (CMRR) of the system, and at the same time reduce the power consumption, volume and cost of the system as much as possible. EEG signals have low frequency and small amplitude, and are easily affected by various noises and interferences. Therefore, when designing the analog front-end circuit, the following conditions need to be met in order to accurately extract the signal:
(1) Variable gain design. A first-stage variable gain amplifier circuit is designed to avoid saturation of the output signal due to changes in EEG signals.
(2) High CMRR and input impedance. The 50 Hz power frequency interference in AC will be introduced, which is a common mode interference for the analog front-end system. Therefore, the circuit should improve the common-mode rejection ratio to suppress common-mode interference. In the skin-electrode interface, there is an interface impedance, which usually reaches the megohm level. Therefore, the input impedance of the analog front-end system must be high enough, otherwise the signal reaching the analog front-end will be seriously attenuated.
(3) Low-input referred noise. Noise performance is one of the most important indicators to be considered in the design of an EEG signal recording system. The input referred noise determines the minimum amplitude of the EEG signal that the system can record. In the design of amplifier circuits, the signal-to-noise ratio is usually used to determine the noise performance that the circuit needs to achieve. The SNR of the front-end circuit of the EEG signal recording system is required to be greater than 40 dB. For the signal recording system, the typical value of the intracranial EEG signal is 500 µVrms, so the equivalent input noise of the analog front-end circuit must be less than 5 µVrms.
(4) Low power consumption. In recent years, wearable medical devices have gradually become a development trend. Such devices require long-term, real-time monitoring and the detection of patients. Therefore, low power consumption design is required to prolong battery life. Excessive power consumption of the EEG recording circuit system will seriously heat up the implanted device, thereby increasing the temperature of the contacted brain tissue, thereby affecting normal data recording, and even causing damage to the brain tissue, which is not allowed. Low power consumption design is one of the technical bottlenecks restricting the implementation of implantable systems. Especially in the case of a large number of recording channels, the power consumption requirements for the single-channel front-end amplifier are very high, usually in the order of µW.
Traditional EEG acquisition equipment usually includes the following main components: an electrode lead interface, preprocessing circuit, analog-to-digital conversion circuit (ADC), processor circuit, storage module, communication circuit and power supply module. Some equipment will add some expansion modules for specific research, such as sensors for collecting other information[
Figure 2.(Color online) Block diagram of a typical EEG acquisition system.
Electrode
The recording electrode is a sensor that converts the brain electrical activity signal of the human body into a voltage (current) signal that can be processed by the circuit. It is a bridge connecting the electrical activity of brain neurons and the circuit system. Electrodes are mainly divided into non-invasive electrodes and invasive electrodes. Non-invasive electrodes are generally used to obtain EEG signals on the surface of the scalp, mainly including wet electrodes, dry electrodes and non-contact electrodes: Wet electrodes are generally composed of conductive gel, Ag/AgCl[
Invasive electrodes, also known as implantable electrodes, generally refer to that which is surgically implanted on the surface of the dura mater or cortex, or which penetrates the brain to measure and record EEG signals from individual cells or cell groups. Invasive electrodes fall into two broad categories: penetrating and non-penetrating cortical electrodes. Needle-penetrating cortical electrodes are defined as individual microfilaments, microfilament bundles, or arrays that are used for precise positioning by the recording site in the brain, which is used to measure action potentials or LFP[
Preprocessing circuit
After the signal is fed through the electrode interface, it is transmitted to the analog channel. The amplitude of EEG is only at the μV level. During the signal acquisition process, it is extremely susceptible to various interferences from the human body itself and the external environment. The analog front-end circuit needs to remove these interferences as much as possible while amplifying the signal, otherwise the quality of the collected signal will be affected. Common interference factors include electrode direct current (DC) imbalance[
The main function of the preprocessing circuit is to amplify and filter the original EEG signal. It is the key factor to determine the final signal quality of the EEG acquisition equipment, so it is the crucial research field of the EEG acquisition circuit part. The preprocessing circuit is mainly composed of a low-noise instrumentation amplifier (IA), a programmable gain amplifier (PGA) and a filter. The main function is to provide appropriate gain to the EEG signal while filtering out interference and noise to ensure the quality of the signal input to the ADC. Among them, the IA amplifies the EEG signals collected by the electrodes, and suppresses interference such as electrode offset voltage and low-frequency noise; the PGA selects the corresponding amplification factor according to the amplitude of the EEG signals, and adjusts the swing of the output signal. The signal-to-noise ratio is optimized; the filter filters out the frequency components outside the signal band to avoid noise aliasing in the ADC sampling process.
Instrumentation amplifier (IA)
The IA is the first-stage amplification circuit of the AFE, which provides high gain for the input signal and can attenuate the noise equivalent to the input terminal of the subsequent circuit. Therefore, in the overall architecture of the front-end circuit, usually only the low-noise design of the IA needs to be considered. However, considering the existence of offset and common-mode interference, too high gain will easily cause the output of the amplifier to be saturated. The gain of the IA is generally set at 40 dB. At the same time, the IA must have a large CMRR to suppress 50/60 Hz power line interference. Also, the IA is directly connected to the electrode, so the input impedance needs to be large enough[
(1) Three-op-amp IA: The three-op-amp structure is the most common structure, usually composed of three op-amps, as shown in
Figure 3.(Color online) Three common IA structures.
(2) AC capacitive coupling IA: As shown in
(3) Current feedback IA: The structure of the current feedback IA is shown in
Programmable gain amplifier (PGA)
Changes in the recording site, environment, and sampling bioelectrodes will cause a wide range of changes in the amplitude of the scalp EEG signal. Therefore, a PGA is required to adjust the overall gain of the pre-circuit to avoid saturation of the output signal. Considering factors such as chip area, noise performance, and circuit power consumption for the first-stage amplifier, the gain of the preamplifier may not meet the amplification requirements of the EEG signal, so a second-stage amplifier is needed to re-amplify the pre-stage signal. The second stage amplifier needs to be designed with adjustable gain. A PGA is usually designed as a closed-loop circuit structure, and the gain of the amplifier is adjusted through a logic control circuit. Adopting a closed-loop structure, the gain accuracy of the amplifier is high, the linearity is relatively good, and the output range is also large. Therefore, it is very suitable for the front-end circuit for extracting biomedical signals such as EEG. There are currently two mainstream PGAs: variable resistor feedback PGA and variable capacitor feedback PGA. The structure of the resistive feedback PGA is shown in
Figure 4.(Color online) Two common PGA structures.
Filter
Three types of filters are commonly used in EEG acquisition equipment: a high-pass filter, power frequency notch filter and low-pass filter.
At the forefront is the high-pass filter, whose main function is to isolate the DC component of the brain potential. It is generally believed that the frequency components of scalp EEG signals related to cognitive tasks are located at 0.5−100 Hz. The starting frequency of the high-pass filter is generally set at about 0.5 Hz, and is usually realized by an active filter. The band-pass amplitude and cut-off frequency of the active filter will not change due to different loads, and it can also dynamically compensate reactive power[
The main function of the power frequency notch filter is to filter out the 50 Hz power frequency noise mixed in the EEG signal. In the traditional notch design, the notch with a double-T structure is the most commonly used, but this kind of notch has a practical defect, that is, it requires high precision of components. In order to achieve the ideal effect, the accuracy of resistors and capacitors is required to reach 0.1%, which is very difficult for the current capacitor-processing technology[
The role of the low-pass filter is to filter out environmental high-frequency noise interference, mainly environmental radio waves, power supply ripple, and noise caused by active devices themselves[
(1) RC filter: Passive RC filters are generally implemented off-chip, which is not considered in this article. The active RC filter adds a high-gain open-loop operational amplifier.
Figure 5.(Color online) Three common filter structures.
(2) Switched capacitor filter: In order to solve the area problem of the RC filter, the switched capacitor filter uses a set of switches and capacitors to be equivalent to a large resistor, and its resistance value is the ratio of the voltage change value on the capacitor to the average current in one clock cycle.
(3) Gm-C filter: Unlike active RC filters, Gm-C filter circuits only contain capacitors and transconductance units and do not require the use of resistors. The function of the transconductance unit is to convert the input voltage into the current. In the Gm-C filter, the basic unit of the integrator is composed of a transconductance unit and a capacitor, which are then cascaded to form a filter.
Analog-to-digital conversion circuit (ADC)
The analog-to-digital conversion circuit in the acquisition module is a key part of EEG signal acquisition. EEG signals are extremely weak and are easily disturbed by noise. Good ADC performance can prevent EEG signals from being distorted due to noise coupling and improve signal-to-noise ratio, ensuring high-fidelity transmission. At the same time, the effective frequency range of the EEG signal on the surface of the scalp is 0.1−100 Hz. According to the Nyquist sampling theorem, the ADC sampling frequency is at least twice the maximum signal frequency to ensure that no aliasing occurs within the signal bandwidth. In addition, since the DC interference is not filtered out in the ADC front-end circuit, the signal for analog-to-digital conversion will be coupled into the DC component, and the minimum amplitude of the EEG signal is greater than 10 μV. In order to prevent the introduction of excessive quantization noise, the minimum resolution of the ADC should be below 1 μV. In order to ensure the quality of the acquired signal, the ADC generally chooses a device with low noise and large input impedance[
Processor module
According to different functions and applications, some EEG acquisition devices require processors to run signal-processing algorithms. The processor circuits of such devices are usually implemented by DSP, FPGA or microcomputers with high computing power. The processor sends the processed data to the upper computer through the communication interface, or directly outputs to the display screen for display, or directly controls some controlled equipment. However, some EEG devices only need the processor to execute control instructions, and send the collected EEG signal data to the host computer through the communication circuit, and the host computer runs the EEG signal processing algorithm. In order to facilitate the data management in the process, the EEG acquisition equipment usually has a storage module, and the storage space varies according to the application[
Integrated chip
The bulky EEG acquisition equipment in the past was mainly used for scientific research and medical treatment, and it was difficult to apply it in daily life. Since the 1990s, foreign countries have begun to study integrated chips for EEG acquisition[
Qian et al. proposed a micropower low-noise neural front-end circuit capable of recording epileptic fast ripples (FR). The front-end circuit consists of a preamplifier and a 6th-order bandpass filter, which is designed for signal sensing in deep brain stimulators for epilepsy in the future. To improve the noise and power trade-off in the preamplifier, a current-splitting technique combining with an output-branch current scaling technique in a folded-cascode amplifier structure is proposed. Experimental results show that the preamplifier owns 39.4 dB DC gain, 0.36 to 1.3 kHz of −3 dB bandwidth, and 3.07 μVrms total input-referred noise and consume 2.4 μW under a 2.8 V power supply[
Robinet et al. proposed a 32-channel recording ASIC that can realize low-noise amplification and analog filtering functions. Meanwhile, it can also own a 12-bit ADC function and offer programmable output rates by using a serial peripheralinterface (SPI). The entire system is designed for use in a remotely powered wireless implantable ECoG recording system. The measured input-referred noise of each recording channel is 0.7 μVrms on a 0.5−300 Hz bandwidth. The ASIC was implemented in a 0.35-μm CMOS process and the total die area is 86 mm2, the analog power consumption is limited to 134 μW for per channel[
Zhou et al. proposed a mixed-signal AFE ASIC for an EEG acquisition system. It will be attached to the scalp via electrodes and will be used to detect electrical signals from the brain, amplify the collected signals and convert them into digital data. The EEG signals of each channel are amplified by a pre-amplifier, whose gain is from 10 to 1000. After the pre-amplifier, a two-stage low pass filter is used with an optional bandwidth of 80 or 800 Hz. The input referred noise of the whole system is 2.2 μVrms for a signal bandwidth of 1−1 kHz. The ADC has a resolution of 12 bits based on the successive approximation algorithm. The power consumption of the analog part for the ASIC is 3.5 mW under a 1.8 V supply[
Figure 6.(Color online) Block diagram of the EEG acquisition AFE[
Yoo et al. proposed a scalable 8-channel EEG acquisition SoC for the continuous detection and recording of patient-specific seizure activity from scalp EEG. The chip integrates an 8-channel AFE, an 8-channel feature extraction processor, a classification processor and 64 kB of memory. The AFE module is powered by 1.8 V and consumes a total of 66 μW in the 8-channel operation mode. The bandwidth of 30/100 Hz is scalable, and the input reference noise is only 0.91 μVrms in the bandwidth of 0.5−100 Hz. The GBW controller is used to provide real-time gain and bandwidth feedback to the AFE to maintain accuracy[
Xu et al. proposed an 8-channel gel-free EEG/electrode-tissue impedance (ETI) acquisition system, which consisted of nine active electrodes (AEs) and one back-end (BE) analog signal processor. The AEs are applied to amplify the weak EEG signals, and their low output impedance can suppress artifacts and 50/60 Hz power frequency interference. The common-mode feed-forward (CMFF) scheme improves the CMRR of the AE pairs by 25 dB. The BE is applied to post-processes and digitize the analog outputs of the AEs, which can also configure them by a single-wire pulse width modulation (PWM) protocol. The input impedance of each EEG channel is 1.2 GΩ at 20 Hz, and 1.75 µVrms (0.5−100 Hz) input-referred noise, 84 dB CMRR and ±250 mV electrode offset rejection capability. The EEG acquisition system was fabricated in a standard 0.18 µm CMOS process, and the power is less than 700 µW under a 1.8 V supply[
Muller et al. proposed a minimally invasive 64-channel ECoG signal acquisition system. The device consists of a highly flexible, high-density 64-channel electrode array and a flexible antenna with a total chip power consumption of only 225 μW, enabling long-term stable neural recordings. The input reference noise of the AFE module can reach as low as 1.23 μVrms at a chopping frequency of 16 kHz, but the system has not been tested in a hospital environment and is slightly deficient compared to the ECoG signals acquired under conventional conditions[
Smith et al. proposed a technique for AFE design specific to EEG. This paper improves noise performance of a similar system and proposes an equalization technique, which can reduce the ADC dynamic range requirements and does not require a variable gain amplifier (VGA). The prototype is fabricated in 1p9m 65 nm CMOS, which can take advantage of the presented findings to realize high-fidelity, full-spectrum ECoG recording. The power consumption of the entire analog front end is 1.08 μW over a 150 Hz bandwidth and only 7 bits of ADC resolution[
Tohidi et al. proposed a low-power instrumentational amplifier (IA), which is designed for EEG signal acquisition targeting seizure detection. The power of the proposed structure per channel is 0.92 µW under a 0.8 V supply. Due to the use of buffer structure and impedance-enhanced loop, the input impedance reaches 160 and 16 GΩ at 1 and 10 Hz, respectively. Also, the chopping techniques lead to an input-referred noise of 1.7 µVrms over the bandwidth of 0.5−100 Hz with a noise efficiency factor (NEF) of 3.87 and a CMRR of 137 dB[
Figure 7.(Color online) Proposed instrumentational amplifier structure[
Karimi-Bidhendi et al. proposed two ultra-low-power EEG signal acquisition analog front-end systems working in the weak inversion area. The power supply voltage of the two AFEs is only 0.4/0.6 V, and a capacitively coupled P−N complementary OTA is used to achieve power consumption of 0.216/0.69 μW and equivalent input noise of 2.19/2.3 μVrms respectively. Human tests have shown that the system can reliably record EEG and ECoG neural signals and can be used as the basis for a small implantable ultra-low-power EEG signal acquisition unit, but it still needs to further reduce the sensitivity of the front-end to environmental noise[
Wu et al. proposed a 16-channel AFE EEG signal acquisition device for a closed-loop seizure control system. It is consisted of 16 input protection circuits, 16 auto-reset chopper-stabilized capacitive-coupled instrumentation amplifiers (AR-CSCCIA) with band-pass filters, 16 programmable transconductance gain amplifiers, a multiplexer, a transimpedance amplifier, and a SAR ADC. The proposed AFE amplifier has 49.1/59.4/67.9-dB programmable gain and 2.02-µVrms input-referred noise in a bandwidth of 0.59−117 Hz. The power consumption of the proposed AFE amplifier is 3.26 µW of per channel, and the NEF is 3.36. The system has been successfully performed in vivo animal tests and verifies its functions. Experimental results show that the proposed AFE circuit is suitable for closed-loop seizure control devices[
Figure 8.(Color online) System architecture of the proposed ECoG-signal acquisition system[
Tohidi et al. proposed a low-power-consumption, high-input impedance AFE circuit design scheme for a continuous working EEG acquisition system. In addition to low-voltage low-power techniques, the design uses active electrodes to increase input impedance and a right-leg drive circuit to improve the common-mode rejection ratio (CMRR). After measurement, the input impedance is 102 GΩ@1 Hz and 5.2 GΩ@20 Hz, the equivalent input noise in the 0.5 Hz 1.2 kHz integration band is 1.5 μVrms, and the CMRR is greater than 108 dB[
Tang et al. proposed a multi-channel AFE ASIC for wearable EEG recording systems. Chopping stabilization (CS) and time-division-multiplexing (TDM) are combined uniformly to optimize the input-referred noise and the system-level CMRR targeting multi-channel AFE. Using this TDM/CS structure, multiple channels can share the same secondary amplifier, helping to reduce chip size and power consumption. The system includes dual feedback loops for input impedance boosting and electrode offset cancellation. The power consumption of the AFE is 24 µW at a 1 V supply voltage. The input referred noise is 0.63 µVrms in 0.5−100 Hz and the input impedance is boosted to 560 MΩ at 50 Hz. The CMRR of the amplifier and system-level AFE are 89 and 82 dB, respectively[
Gao et al. proposed a two-stage time-division multiplexing analog front-end (TDMAFE) for multi-channel EEG signal acquisition. They proposed a four-channel synchronous TDM technology eliminating the mismatch among the four-channel signals. To increase input impedance, an impedance-boosted chopper based on the differential difference amplifier is proposed. And to eliminate DC offset, an improved low-noise digitally controlled DC servo loop (DCDSL) is proposed. The entire AFE system is 1.2 mm2 and the power consumption is 30.6 μW under a 1.2-V supply. Post-simulation results show that the gain in the AFE ranges from 34 to 60 dB. The range of the input offset cancellation is ±300 mV. The equivalent input impedance is boosted to 1.6 GΩ. The equivalent integrated noise is 1.66 μVrms at the range of 0.1−100 Hz[
Huang et al. proposed an AFE amplifier (AFEA), which can be applied for EEG recording. A capacitively coupled chopper instrumentation amplifier (CCCIA) with chopper modulation is used in the proposed AFEA to suppress flicker noise within the bandwidth and achieve lower noise. In this paper, the EEG AFEA has the tunable gains of 60.4/69.6/79.2 dB. The input referred noise within the bandwidth is 0.839 μVrms, the area is 2.576 mm2 and power consumption is 83.2 μW[
Epilepsy signal classification technology
With the advancement of artificial intelligence technology in recent years, researchers continue to dig in the field of automatic diagnosis of epilepsy based on EEG, and have obtained many detection methods to solve the problem. This algorithm is of great help to the automatic detection of epilepsy, because it can independently learn the classification rules of the data and improve the performance of data expansion. The core issues of epilepsy detection technology are feature selection and classifier design. Feature selection aims to select useful features depicting epileptic signals; the classification stage classifies the signals according to the selected features, showing different classification results of epileptic states. The following is an introduction to the research progress of epilepsy signal classification technology from four aspects: several public datasets currently used in epilepsy detection, feature extraction and the selection of epilepsy EEG signals, an epilepsy detection algorithm based on classical machine learning and an epilepsy detection algorithm based on deep learning.
Publicly available epilepsy detection datasets
It is necessary for a lot of scientists and researchers to use a suitable dataset for assessing the performance of their proposed methods. Therefore, we need to collect a large number of EEG signals for the detection and prediction of epileptic seizures. Among the commonly used methods for detecting brain activity, EEG recording is the most popular. These records play a key role in exploring the development of machine learning classifiers, which can help to find effective methods for epilepsy seizure detection in different aspects. The existence of publicly available datasets is important because they provide a reference base to analyze experimental results and allow researchers to compare experimental results with others. In this section, we introduce several publicly available datasets that are popular in this research field.
Children Hospital Boston, Massachusetts Institute of Technology—EEG dataset
This dataset was collected in Boston Children’s Hospital, which included EEG recordings of pediatric patients with intractable seizures. Physicians monitored the participants for several days after antiepileptic drugs were stopped, to describe their seizures state and evaluate their suitability for surgical intervention.
The entire dataset includes 23 cases from 22 subjects (5 males, 3 to 22 years; 17 females, 1.5 to 19 years) (the chb21 and the chb01 were collected from the same female subject with the interval of 1.5 years)[
Long-term SWEC-ETHZ—EEG dataset
This dataset includes 18 patients of the Inselspital Bern, P1–P18. They were from the epilepsy surgery program[
Bern-Barcelona—EEG dataset
This dataset contains intracranial EEG recordings of five epileptic patients, divided into two categories. In the first category, EEG signals were collected in the epileptic area, which is called focal (F). In the second category, EEG signals were collected in the non-epileptic area which is called non-focal (N). Each category includes the 3750 pairs of recordings. The two signals of F class were collected from the epileptic signal originating channel and another adjacent channel. The signals of the N class were collected from two adjacent channels located in non-epileptic regions. All EEG recordings were filtered by using a 4th-order Butterworth filter, and the filter frequency band is from 0.5 to 150 Hz. All signals of the original dataset were recorded lasting 20 s, where the sampling frequency is 1024 Hz. Then, all signals were down-sampled to 512 Hz. At last, the median was subtracted across all channels. The original dataset did not include recordings of the epileptic seizures and recordings of 3 h after the last seizure[
Bonn University—EEG dataset
This dataset includes 500 recordings of 500 different patients, where each record lasts 23 s[
The Freiburg—EEG dataset
This dataset includes 21 patients’ invasive EEG recordings, which with medically intractable focal epilepsy. These data were collected in the Epilepsy Center of the University Hospital of Freiburg, Germany[
Feature extraction and selection
It is an important step to realize the automatic detection of epilepsy by analyzing the signal and extracting effective features as the classification basis. Reasonable and typical epileptic EEG features can fully characterize the EEG signal pattern, and effectively describe the difference between EEG signals in various states such as seizures and normal states, highlighting the difference between spikes and background signals, thereby helping the classification model to identify epileptic seizures effective screening. The quality of features seriously affects the final classification performance.
In general, the features used for epilepsy detection can be divided into the following four categories: time-domain features, time-frequency domain features, and nonlinear features.
Time-domain features are the most basic features in EEG signal processing. Groups of statistical parameters have been frequently used to discriminate between ictal and normal patterns, because it is assumed that EEG statistical distributions during a seizure and normal periods are different. These parameters are mean, variance, mode, median, skewness, and kurtosis. The minimum and maximum values are also used to quantify the range of data or the magnitude of the signal baseline. Other statistical parameters include coefficient of variation (CV) defined as the ratio of the standard deviation (SD) to the sample mean that explains the dispersion of the data in relation to the population mean. Energy, average power, and root mean squared value (RMS) are mutually relevant to amplitude measurements. The energy is a summation of a squared signal, the average power is the signal mean square, and the RMS is the square root of the average power[
Frequency domain analysis is based on the assumption that EEG signals have stationary characteristics. The EEG signal contains a variety of frequency components. By transforming the EEG signal from the time domain to the frequency domain, the information of different frequencies is analyzed and features are extracted. For epileptic EEG signals, compared with the interictal period, the ictal EEG will have obvious characteristic waves such as sharp waves and spikes, and the signal frequency will change significantly. The frequency domain features of EEG signals include power spectral density, high-order spectrum, differential entropy, etc.[
If the amount of information contained in the features is considered, neither the pure time-domain features nor the frequency-domain features can completely describe an EEG signal, and the EEG analysis based on the assumption of stationarity is not rigorous. Therefore, the researchers turned their attention to the time-frequency analysis method, and re-expressed the non-stationary EEG signal and extracted the corresponding features through time-frequency transformation and other means. Time-frequency analysis is mainly connected by short-time Fourier transform at the beginning, but because the size and shape of its window are fixed, it cannot be well adapted to EEG signals with fast frequency and time changes. The wavelet transform adaptively changes the shape of the time window through stretching and translation, so wavelet analysis is a representative method of time-frequency analysis. In addition to wavelet analysis, commonly used time-frequency domain analysis methods include wavelet packet decomposition, Gabor transform and so on[
Methods such as time domain, frequency domain or time-frequency analysis are mainly based on the theoretical analysis of linear systems. Since EEG signals have obvious nonlinear characteristics, processing methods for nonlinear signals are needed. Therefore, nonlinear dynamics theory is widely used in the field of EEG signal processing. Nowadays, nonlinear dynamic indicators such as approximate entropy, spectral entropy and sample entropy, etc. are commonly used in EEG signal analysis[
Epilepsy detection algorithm based on classical machine learning
In the epilepsy detection algorithms based on classical machine learning, data preprocessing of EEG signals, effective feature extraction of processed signals, and classification based on extracted features are the main tasks in the detection algorithm. Researchers have achieved automatic detection and the classification of epileptic signals by using time-frequency domain methods[
Figure 9.(Color online) A typical random forest for epileptic seizure detection.
Gotman et al.[
Classical machine learning models mainly design effective features artificially based on signals. This stage requires researchers to have professional knowledge related to epileptic EEG signals. Although the automatic seizure detection technology based on classical machine learning has shown good accuracy using traditional signal processing (signal processing, SP) and machine learning techniques, the detection and classification of epileptic EEG signals is a difficult task. Accurately classified epileptic vs. non-ictal cases performed poorly on the three-category task of normal, ictal, and interictal[
Epilepsy detection algorithm based on deep learning
In recent years, with the rapid development of deep learning (DL), deep neural network (DNN) models have been widely used in various fields. As a class of machine learning methods, deep learning can automatically encode the hierarchical structure of data-independent elements and adapt to the internal structure of the data, thereby extracting deep features that are not easy to be observed and extracted. Compared with traditional feature extraction models, deep learning methods eliminate the dependence on manually extracted features.
Figure 10.(Color online) A typical 2D-CNN for epileptic seizure detection.
Convolutional neural network (CNN)-like models are capable of extracting features from input data. With the continuous research of deep learning theory by researchers and the improvement of numerical computing hardware equipment, CNN has developed rapidly in recent years, and various CNN models such as AlexNet, VGG and 3D network[
A recurrent neural network (RNN) focuses on the correlation of outputs at different times in the sequence. Among them, a long-short-term memory network (LSTM) is the most widely used model in RNN, which can fully demonstrate its superior performance in the processing of time series problems. Hu et al.[
Since the automatic detection of epileptic EEG signals has a significant effect on the clinical diagnosis of epilepsy, it is necessary to study automatic detection based on various models.
Epilepsy detection hardware implementation
At present, researchers have developed many complete systems for detecting and predicting epilepsy. These systems have the advantages of high accuracy and energy saving, which provide ideas and hope for the development of wearable epilepsy monitoring equipment.
Yoo et al. proposed an ultra-low power scalable EEG acquisition system on Chip (SoC) for continuous seizure detection and recording, with a fully integrated patient-specific Support Vector Machine (SVM)-based classification processor. With the SoC, a small form factor, patch-type long-term seizure monitoring and recording device can be realized. The SoC integrates eight-channel, bandwidth and gain-scalable AFE with low noise, low power instrumentation circuits, an ADC, eight-channel classification processor, and storage on a single 25 mm2 chip (0.18 μm 1P6M standard CMOS). The seizure detection processor is tested with the CHB-MIT database, and the SoC is verified with a rapid eye blink test, which shows typical accuracy of 84.4% with 2.03 μJ/classification energy efficiency[
Altaf et al. proposed a 16-channel noninvasive closed-loop beginning- and end-of-seizure detection SoC. The dual-channel charge recycled (DCCR) analog front end (AFE) achieves chopping and time-multiplexing an amplifier between two channels simultaneously which exploits fast-settling DC servo-loop with current consumption and NEF of 0.9 µA/channel and 3.29/channel, respectively. The dual-detector architecture (D2A) classification processor utilizes two linear support-vector machine (LSVM) classifiers based on digital hysteresis to enhance both the sensitivity and the specificity simultaneously. The pulsating voltage transcranial electrical stimulator (PVTES) automatically configures the number of pulses to control the amount of charge delivered based on skin-electrode impedance variation in efforts to suppress the seizure activity, while burning only 2.45 µW. The 25 mm2 SoC implemented in 0.18 µm CMOS consumes 2.73 µJ/classification for 16 channels with an average sensitivity, specificity, and latency of 95.7%, 98%, and 1 s, respectively[
Altaf et al. proposed an eight-channel patient specific scalable EEG acquisition SoC to continuously detect the patient-specific seizure electrical onset. A non-linear machine learning algorithm is employed using a 0.18 μm 1P6M standard CMOS process with an area of 25 mm2. The SoC integrates the multi-channel analog front-end with low-noise, low-power instrumentation circuits, an ADC, a classification processor, and an SRAM. The seizure detection processor is tested with CHB-MIT database, and the SoC is verified with rapid eye blink tests, which shows the sensitivity of 95.1% and 0.27 false alarm/hour while consuming 1.83 μJ/classification[
Lin et al. proposed a smart headband for epileptic seizure detection. The proposed headband consists of four key components: 1) an analog front-end circuitry; 2) an epileptic seizure detection tag (ESDT); 3) a bluetooth low-power chip; and 4) customized electrodes. The overall system is fully integrated into a fabric headband weighting only 50.3 g and dissipates 55.89 mW. This compact smart headband with a corresponding APP for epileptic seizure detection can mprove the patients’ quality of life by monitoring their health onditions at anytime and anywhere. The epileptic seizure detection algorithm inside ESDT is validated by using Boston Children’s Hospital’s CHB-MIT scalp EEG clinical database with the detection rate of 92.68% and the false alarm of 0.527 h−1[
Huang et al. proposed the first support vector machine (SVM) processor that supports on-chip active learning for seizure detection. Alternating direction method of multipliers (ADMM), minimum-redundancy maximum-relevance (mRMR), and low-rank approximation are utilized to reduce the computational complexity and memory storage by 99.4% and 90.4%, respectively. Hardware complexity is reduced by 87% through folded architecture with reconfigurable processing elements. The 4.5 mm2 SoC implemented in 40 nm CMOS, achieving the best detection performance with a 96.1% accuracy and a 0.34% false alarm rate in 0.71 s with 1.9 mW[
Liu et al. proposed a reconfigurable biomedical AI processor with adaptive learning. It is mainly composed of five modules: a reconfigurable neural-network engine (RNNE), various biomedical-processing engines (BPEs), an adaptive-learning engine (ALE), a reconfigurable FIR engine (RFE) and an adaptive biomedical signal compression engine (ABSCE). For an EEG seizure detection task using the Bonn database, two-stage event-driven AI processing is enabled. Compared to non-event-driven AI processing, the power consumption can be significantly reduced by 49% with accuracy of 99.84%. The 1.74 mm2 SoC implemented in 60 nm CMOS consumes 2.06 µJ/classification[
Zhang et al. proposed a fully programmable patient-specific closed-loop epilepsy tracking and suppression SoC. The proposed two-cycle analog front end (2C-AFE) obtains a 9.8-b effective number of bits (ENOB) with 8 × capacitive digital-to-analog converter (CAPDAC) area reduction and 4 × switching energy saving compared to a conventional 10-b SAR with an identical unit capacitor size. The entire SoC with 16 surface EEG recording channels consumes an ultra-low energy of 0.97 µJ/class and occupies a miniaturized area of 0.13 mm2/ch in 40 nm CMOS, achieving real-time concurrent seizure detection and raw EEG recording. Verified with the CHB-MIT database, the guided time-channel averaging (GTCA) neural processor achieves the vector-based sensitivity, the specificity, and the latency of 97.8%, 99.5%, and <1 s, respectively[
Zhu et al. proposed a model for seizure detection based on 1D-CNN, which loses only a small amount of accuracy but greatly reduces computational complexity and model size. Furthermore, a low power consumption and high efficiency based on 1D-CNN hardware accelerator architecture is proposed. It only takes 170 μs to perform a round of inference on Xilinx ZC706 evaluation platform at 200 MHz, which is much more efficient than other hardware network accelerators[
Chua et al. proposed SOUL: Stochastic-gradientdescent-based Online Unsupervised Logistic regression classifier. After an initial offline training phase, continuous online unsupervised classifier updates are applied in situ, which improves sensitivity in patients with drifting seizure features. SOUL was tested on two human EEG datasets: the Children’s Hospital Boston and the Massachusetts Institute of Technology (CHB-MIT) scalp EEG dataset and a long (>100 h) intracranial EEG dataset. It was able to achieve an average sensitivity of 97.5% and 97.9% for the two datasets, respectively, at >95% specificity. Sensitivity improved by at most 8.2% on long-term data when compared to a typical seizure detection classifier. SOUL was fabricated in Taiwan Semiconductor Manufacturing Company (TSMC’s) 28 nm process occupying 0.1 mm2 and achieves 1.5 nJ/classification energy efficiency, which is at least 24 × more efficient than state-of-the-art[
Li et al. proposed a novel low-latency parallel Convolutional Neural Network (CNN) architecture that has between 2-2800 × fewer network parameters compared to state-of-the-art (SOTA) CNN architectures and achieves 5-fold cross validation accuracy of 99.84% for epileptic seizure detection, and 99.01% and 97.54% for epileptic seizure prediction, when evaluated using the University of Bonn EEG, CHB-MIT and SWEC-ETHZ seizure datasets, respectively. They investigated the effects of non-idealities on our system and investigate quantization aware training (QAT) to mitigate the performance degradation due to low analog-to-digital converter (ADC)/digital-to-analog converter (DAC) resolution. Finally, we propose a stuck weight offsetting methodology to mitigate performance degradation due to stuck RON/ROFF memristor weights, recovering up to 32% accuracy, without requiring retraining. The CNN component of their platform is estimated to consume approximately 2.791 W of power while occupying an area of 31.255 mm2 in a 22 nm FDSOI CMOS process[
Conclusion and outlook
Automatic detection of epilepsy, that is, automatic identification of epileptic seizures, can greatly reduce the burden on medical workers and reduce the uncertainty of doctors' subjective judgment of the disease. Since EEG is a low-cost, easy-to-acquire and non-invasive clinical physiological signal, EEG-based automatic detection of epilepsy is an effective means to improve the efficiency of epilepsy detection and treatment. This paper first systematically expounds the process of automatic detection of epilepsy based on EEG, and conducts a detailed and comprehensive investigation and summary of the specific methods involved in each step.
According to the above research and analysis, combined with the actual application scenario requirements of EEG-based automatic epilepsy detection, this paper looks forward to the future research direction from the following aspects:
(1) The low-power design of wearable devices is the general trend. It is necessary to study circuit design schemes suitable for different processes at lower power supply voltages, and reduce the demand for low-pass filters by optimizing the bandwidth of the front-end amplifier, thereby saving power consumption. Hardware performance such as common-mode rejection ratio, analog-to-digital conversion accuracy and speed, and dynamic range is the guarantee for the accuracy of EEG acquisition signals and the basis for subsequent EEG signal processing and criteria. For dry electrodes and non-contact electrode systems, motion artifacts will seriously affect the normal operation of the AFE circuit, and it is necessary to study the analog circuit suppression method for motion artifacts combined with electrode impedance detection technology.
(2) When using traditional machine learning methods, extracting effective classification features is the key to epilepsy detection. Seizure detection by traditional methods has achieved a high level of detection, but how to distinguish epilepsy subtypes, how to distinguish between onset and interictal periods to realize the prediction of epilepsy, is still a problem worth exploring.
(3) In recent years, with the rapid development of deep learning, a large number of automatic detection methods based on deep learning have been widely used in automatic detection of epilepsy, and achieved good results. However, this method also has many challenges. a) Deep learning often relies on a large amount of data, but limited by the data set in actual research, it is difficult to obtain a large number of training samples, which will have a great impact on the accuracy and robustness of the model. b) The existing public datasets are almost all EEG signal fragments, which are different from the continuous real-time signals in the actual scene. When solving practical problems, the neural network model trained with EEG fragment signals may not be well adapted to the real scene. c) Due to the limited computing resources in actual use, lightweight neural network models will be more practical.
(4) With the continuous development of wearable devices, in addition to the existing feature extraction for EEG signals, the fusion of other physiological features also has considerable research value. Previous studies have shown that blood oxygen saturation is related to the termination of epileptic seizures[
References
[2] J R Stevens. Seizure occurrence and interspike interval. Arch Neurol, 26, 409(1972).
[24] T J Sullivan, S R Deiss, G Cauwenberghs. A low-noise, non-contact EEG/ECG sensor, 154(2008).
[29] J G Webster, J W Clark. Medical instrumentation: application and design. John Wiley & Sons(2010).
[30] J C Huhta, J G Webster. 60-HZ interference in electrocardiography. IEEE Trans Biomed Eng, 20, 91(1973).
[39] A Bakker, J Huijsing. High-accuracy CMOS smart temperature sensors. Springer(2000).
[47] B Babušiak, Š Borik. Bio-Amplifier with programmable gain and adjustable leads, 616(2013).
[49] N Marchon, G Naik. A novel linear phase FIR high pass filter for biomedical signals, 147(2019).
[58] W C Lai. SAR ADC with optical frontend for biosensor applications, 1(2021).
[103] C Mahjoub, R Le Bouquin Jeannès, T Lajnef et al. Epileptic seizure detection on EEG signals using machine learning techniques and advanced preprocessing methods. Biomedical Engineering / Biomedizinische Technik, 65, 1(2019).
[131] A H Shoeb. Application of machine learning to epileptic seizure onset detection and treatment. Massachusetts Institute of Technology, 1, 1(2009).
[148] M Mursalin, S S Islam, M K Noman et al. Epileptic seizure classification using statistical sampling and a novel feature selection algorithm(2019).
[157] S Roy, I Kiral-Kornek, S Harrer. ChronoNet: A deep recurrent neural network for abnormal EEG identification. Conference on Artificial Intelligence in Medicine in Europe. Springer, 47(2019).
[160] R H Gabr, A I Shahin, A A Sharawi et al. A deep learning identification system for different epileptic seizure disease stages. J Eng Appl Sci, 67, 925(2020).