• Chinese Optics Letters
  • Vol. 15, Issue 7, 071402 (2017)
Zhijun Yuan1、2, Haibo Zhang1、2, Ren Ye1、2, Jun Zhou1、2、*, Yunrong Wei1、2, and Qihong Lou1、2
Author Affiliations
  • 1Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai 201800, China
  • 2Shanghai Key Laboratory of All Solid-state Laser and Applied Techniques, Shanghai 201800, China
  • show less
    DOI: 10.3788/COL201715.071402 Cite this Article Set citation alerts
    Zhijun Yuan, Haibo Zhang, Ren Ye, Jun Zhou, Yunrong Wei, Qihong Lou. Wavelength calibration of narrowband ArF laser with iron hollow cathode lamp[J]. Chinese Optics Letters, 2017, 15(7): 071402 Copy Citation Text show less
    Spectrum of the line-narrowed ArF laser (FWHM 0.376 pm).
    Fig. 1. Spectrum of the line-narrowed ArF laser (FWHM 0.376 pm).
    Experimental setup for wavelength calibration of the ArF laser. HCL, hollow cathode lamp; HV, high voltage; PC personal computer; HR, high-reflection mirror.
    Fig. 2. Experimental setup for wavelength calibration of the ArF laser. HCL, hollow cathode lamp; HV, high voltage; PC personal computer; HR, high-reflection mirror.
    Wavelength with the position of the laser spot on the LDA.
    Fig. 3. Wavelength with the position of the laser spot on the LDA.
    Waveform of iron OG signal recorded by an oscilloscope.
    Fig. 4. Waveform of iron OG signal recorded by an oscilloscope.
    OG signal amplitude with discharge current.
    Fig. 5. OG signal amplitude with discharge current.
    OG signal amplitude at a different laser power and discharge current.
    Fig. 6. OG signal amplitude at a different laser power and discharge current.
    OG signal amplitude recorded in the wavelength range of 193.38500–193.39500 nm.
    Fig. 7. OG signal amplitude recorded in the wavelength range of 193.38500–193.39500 nm.
    ParametersValues
    Diffraction order of the grating k90
    Distance between groves d (mm)1/112.96
    Diffraction angle of the grating θ (deg.)79
    Pixel length (μm)25
    Focal length f (mm)200
    Table 1. Key Parameters of the Wavelength Measurement
    Zhijun Yuan, Haibo Zhang, Ren Ye, Jun Zhou, Yunrong Wei, Qihong Lou. Wavelength calibration of narrowband ArF laser with iron hollow cathode lamp[J]. Chinese Optics Letters, 2017, 15(7): 071402
    Download Citation