• Laser & Optoelectronics Progress
  • Vol. 57, Issue 13, 132201 (2020)
Mingchun Feng1、*, Liang Xu2, and Yujie Wang1
Author Affiliations
  • 1School of Mechanical and Electrical Engineering, Chuzhou University, Chuzhou, Anhui 239000, China
  • 2Key Laboratory of Environment Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    DOI: 10.3788/LOP57.132201 Cite this Article Set citation alerts
    Mingchun Feng, Liang Xu, Yujie Wang. Collimation Factors and Performance Analysis Based on Michelson Interferometer[J]. Laser & Optoelectronics Progress, 2020, 57(13): 132201 Copy Citation Text show less
    Michelson interferometer
    Fig. 1. Michelson interferometer
    Influence of tilt of movable mirror
    Fig. 2. Influence of tilt of movable mirror
    Influence of mechanical vibration. (a) Interferogram at 4000 cm-1; (b) partial enlarged view of Fig.(a)
    Fig. 3. Influence of mechanical vibration. (a) Interferogram at 4000 cm-1; (b) partial enlarged view of Fig.(a)
    Schematic diagram of dynamic alignment system
    Fig. 4. Schematic diagram of dynamic alignment system
    Measured of spectrum
    Fig. 5. Measured of spectrum
    100% line of spectrometer. (a) All band; (b) 2500-2600 cm-1 band
    Fig. 6. 100% line of spectrometer. (a) All band; (b) 2500-2600 cm-1 band
    100% line of 1900-2200 cm-1 band
    Fig. 7. 100% line of 1900-2200 cm-1 band
    Measured of resolution
    Fig. 8. Measured of resolution
    Wavenumber accuracy of 906.62 cm-1
    Fig. 9. Wavenumber accuracy of 906.62 cm-1
    Wavenumber accuracy of 1601.15 cm-1and 3081.87 cm-1
    Fig. 10. Wavenumber accuracy of 1601.15 cm-1and 3081.87 cm-1
    Wavenumber repeatability of 906.62 cm-1
    Fig. 11. Wavenumber repeatability of 906.62 cm-1
    Wavenumber repeatability of 1601.15 cm-1 and 3081.87 cm-1
    Fig. 12. Wavenumber repeatability of 1601.15 cm-1 and 3081.87 cm-1
    SerialnumberCharacteristic peakposition/cm-1Measured peakposition/cm-1Error /cm-1
    1906.62906.640.02
    21601.151601.160.01
    33081.873081.85-0.02
    Table 1. Peak position of polystyrene
    Mingchun Feng, Liang Xu, Yujie Wang. Collimation Factors and Performance Analysis Based on Michelson Interferometer[J]. Laser & Optoelectronics Progress, 2020, 57(13): 132201
    Download Citation