    Jingyou Liu, Feng Lei. Measurement of Lens-Center Thickness Based on Low-Coherence Interference with Transmitted Illumination[J]. Laser & Optoelectronics Progress, 2019, 56(12): 121201 Copy Citation Text show less
    Principle diagram for measuring lens-center thickness based on low-coherence interference with transmitted illumination
    Fig. 1. Principle diagram for measuring lens-center thickness based on low-coherence interference with transmitted illumination
    Principle diagram for adjusting optical path difference using pair of wedge prisms
    Fig. 2. Principle diagram for adjusting optical path difference using pair of wedge prisms
    Equal-path interference fringes before and after lens. (a) Before lens; (b) after lens
    Fig. 3. Equal-path interference fringes before and after lens. (a) Before lens; (b) after lens
    [in Chinese]
    Fig. 3. [in Chinese]
    Method for extending measuring range of lens-center thickness
    Fig. 4. Method for extending measuring range of lens-center thickness
    nPnΔx /μmθsinθΔtC /nm
    5°44'0.1±200
    nPn=nP±22°52'0.05±100
    1°26'0.025±50
    6°12'0.10799±200
    1.475901.51390±23°5'0.05399±100
    1°33'0.02700±50
    5°20'0.09260±200
    1.513901.47590±22°39'0.04630±100
    1°20'0.02315±50
    Table 1. Measuring precision under different conditions
    No.Nominal value of tested lensMeasured value of center thickness
    Dimeter ϕ /mmFocal length f' /mmCenter thickness TC /mm(tC±δ) /mmRelative error /%
    525.4-150.03.0±0.12.9286±0.00090.03
    712.7-25.42.0±0.12.0151±0.00080.04
    912.7150.04.0±0.14.0581±0.00090.02
    1225.4100.04.0±0.14.0375±0.00090.02
    1325.4-150.03.0±0.12.7990±0.00110.04
    1425.4400.03.5±0.13.5294±0.00080.02
    1810.0-20.01.5±0.11.4743±0.00100.07
    196.0-9.801.0±0.11.1094±0.00060.05
    Table 2. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 10°0'
    No.Nominal value of tested lensMeasured value of center thickness
    Dimeter ϕ /mmFocal length f' /mmCenter thickness TC /mm(tC±δ) /mmRelative error /%
    525.4-150.03.0±0.12.9227±0.00020.01
    712.7-25.42.0±0.12.0139±0.00090.05
    1325.4-150.03.0±0.12.7980±0.00070.02
    1425.4400.03.5±0.13.5226±0.00090.02
    1810.0-20.01.5±0.11.4778±0.00060.04
    196.0-9.801.0±0.11.1044±0.00050.05
    Table 3. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 5°30'
    No.Nominal value of tested plate or lensMeasurementCorrected measurement
    Dimeterϕ /mmFocal lengthf' /mmCenter thicknessTC /mm(tC±δ) /mmRelativeerror /%(tC±δ) /mmRelativeerror /%
    1--2.2±0.12.2331±0.00080.04--
    2--3.8±0.13.8447±0.00110.03--
    825.41505.0±0.12.8073±0.00090.035.0404±0.00120.02
    1725.4308.1±0.14.2573±0.00110.028.1020±0.00160.02
    Table 4. Examples for expanding measuring range
    Jingyou Liu, Feng Lei. Measurement of Lens-Center Thickness Based on Low-Coherence Interference with Transmitted Illumination[J]. Laser & Optoelectronics Progress, 2019, 56(12): 121201