    DONG Hai-sheng, ZHANG Li-fen, ZHONG Yue, HUANG Jian-ying, CHEN Bin. Feasibility Reaserch on Using Raman Spectroscopy with PLS for the Quantitative Detection of Cholesterol Content in Serum[J]. Spectroscopy and Spectral Analysis, 2013, 33(5): 1253 Copy Citation Text show less

    Abstract

    Taking the serum obtained from daily clinical inspection as study object, the Raman spectra were collected with the range of 250 to 2 400 cm-1. Taking the serum cholesterol content determined by traditional clinical biochemical method as contrast, together with combination of PLS, we established a Raman spectroscopy quantitative models for cholesterol content in serum. The experimental results show that Raman spectroscopy displacement could reflect certain chemical group in organic compound. The correction model of cholesterol content in serum was established with the coefficient R=0.909 6, cross-validation calibration standard deviation (RMSECV) of 0.24, standard error of prediction of test set (RMSEP) was 0.69, and coefficient (R) is 0.926 2. The results showed that application of Raman spectroscopy combined with chemometrics in the rapid noninvasive detection of serum cholesterol content is feasible.
    DONG Hai-sheng, ZHANG Li-fen, ZHONG Yue, HUANG Jian-ying, CHEN Bin. Feasibility Reaserch on Using Raman Spectroscopy with PLS for the Quantitative Detection of Cholesterol Content in Serum[J]. Spectroscopy and Spectral Analysis, 2013, 33(5): 1253