• Spectroscopy and Spectral Analysis
  • Vol. 33, Issue 10, 2869 (2013)
ZHAO Hui-jie*, ZHOU Peng-wei, ZHANG Ying, and LI Chong-chong
Author Affiliations
  • [in Chinese]
  • show less
    DOI: 10.3964/j.issn.1000-0593(2013)10-2869-06 Cite this Article
    ZHAO Hui-jie, ZHOU Peng-wei, ZHANG Ying, LI Chong-chong. Lateral Chromatic Aberrations Correction for AOTF Imaging Spectrometer Based on Doublet Prism[J]. Spectroscopy and Spectral Analysis, 2013, 33(10): 2869 Copy Citation Text show less

    Abstract

    An user defined surface function method was proposed to model the acousto-optic interaction of AOTF based on wave-vector match principle. Assessment experiment result shows that this model can achieve accurate ray trace of AOTF diffracted beam. In addition, AOTF imaging spectrometer presents large residual lateral color when traditional chromatic aberrations correcting method is adopted. In order to reduce lateral chromatic aberrations, a method based on doublet prism is proposed. The optical material and angle of the prism are optimized automatically using global optimization with the help of user defined AOTF surface. Simulation result shows that the proposed method provides AOTF imaging spectrometer with great conveniences, which reduces the lateral chromatic aberration to less than 0.000 3 degrees and improves by one order of magnitude, with spectral image shift effectively corrected.
    ZHAO Hui-jie, ZHOU Peng-wei, ZHANG Ying, LI Chong-chong. Lateral Chromatic Aberrations Correction for AOTF Imaging Spectrometer Based on Doublet Prism[J]. Spectroscopy and Spectral Analysis, 2013, 33(10): 2869
    Download Citation