    Shinan ZHOU, Xiaoyan SHEN, Dongsheng LI, Chenguang WU. Self-calibration Correction Measurement of Micro-angle Based on F-P Etalon Multi-beam Interference Imaging[J]. Acta Photonica Sinica, 2022, 51(4): 0412007 Copy Citation Text show less
    Micro-angle measurement optical path based on F-P etalon[18]
    Fig. 1. Micro-angle measurement optical path based on F-P etalon18
    Self-calibration process schematic diagram of micro-angle measurement system
    Fig. 2. Self-calibration process schematic diagram of micro-angle measurement system
    The algorithm error simulation results of exact fraction method
    Fig. 3. The algorithm error simulation results of exact fraction method
    Simulation results of F-P etalon interval calculation error changing with temperature and humidity
    Fig. 4. Simulation results of F-P etalon interval calculation error changing with temperature and humidity
    Micro-angle measuring device[18]
    Fig. 5. Micro-angle measuring device18
    Concentric interference rings corresponding to different wavelengths
    Fig. 6. Concentric interference rings corresponding to different wavelengths
    λ1 = 546.081 10 nmλ2 = 576.967 84 nmλ3 = 579.074 30 nm
    εj0.766 40.616 30.201 9
    Uεj0.010 20.009 90.009 5
    k0j7 3256 9336 908
    Table 1. Different wavelength λj and its corresponding εj and k0j
    No.δ/μmsδ/μm

    Not self-calibrated

    f1=80 mm

    Upf1)/f1=0.361

    Self-calibrated but not corrected

    f2=78.178 mm

    Upf2)/f2=0.014

    Self-calibrated and corrected

    f '=78.540 mm

    Upf ')/f '=0.007

    α/(″)Upα)/(″)α/(″)Upα)/(″)α'/(″)Upα')/(″)
    0-10.1570.0120.2020.0410.2070.0410.2060.041
    0-20.2690.0120.3470.0410.3540.0410.3530.041
    0-30.2630.0120.3390.0410.3480.0410.3450.041
    1-176.9910.01499.2530.398101.5670.056101.0980.048
    1-277.1690.01399.4830.399101.8020.056101.3310.048
    1-377.2370.01299.5700.399101.8910.056101.4210.048
    2-1153.6550.014198.0850.755202.7010.071201.7660.055
    2-2153.6990.013198.1420.755202.7590.071201.8240.055
    2-3153.8060.013198.2800.756202.9000.071201.9640.055
    3-1230.2240.012296.7941.111303.7100.087302.3090.062
    3-2230.2910.012296.8801.111303.8000.087302.3970.062
    3-3230.4340.013297.0641.112303.9880.087302.5840.062
    4-1306.9010.012395.6411.468404.8630.102402.9930.069
    4-2306.9580.012395.7151.468404.9370.102403.0680.069
    4-3307.0990.011395.8961.469405.1230.102403.2530.069
    5-1383.5630.012494.4681.824505.9930.117503.6570.077
    5-2383.6110.012494.5301.825506.0570.117503.7200.077
    5-3383.8190.012494.7981.826506.3300.117503.9930.077
    6-1460.2290.012593.3002.181607.1280.132604.3250.084
    6-2460.3230.012593.4212.181607.2520.132604.4480.084
    6-3460.2970.012593.3882.181607.2170.132604.4140.084
    Table 2. Comparison of micro-angle measurement results before and after self-calibration