• Chinese Journal of Lasers
  • Vol. 47, Issue 2, 207013 (2020)
Qiu Jianrong, Han Tao, Wang Di, Meng Jia, Liu Zhiyi, and Ding Zhihua*
Author Affiliations
  • State Key Laboratory of Modern Optical Instrumentation, College of Optical Science and Engineering, Zhejiang University, Hangzhou, Zhejiang 310027, China
  • show less
    DOI: 10.3788/CJL202047.0207013 Cite this Article Set citation alerts
    Qiu Jianrong, Han Tao, Wang Di, Meng Jia, Liu Zhiyi, Ding Zhihua. Probes for Endoscopic Optical Coherence Tomography: Minimized Design and Depth of Focus Extension[J]. Chinese Journal of Lasers, 2020, 47(2): 207013 Copy Citation Text show less

    Abstract

    Minimized probe is a common requirement in endoscopic optical coherence tomography (OCT). We introduce the development of mainstream designs based on ball lens, fiber lens, graded index fiber, free form lens and free-lens, summarize their advantages and disadvantages, and put forward some suggestions for miniaturization of the probe. The development of probe with extended depth of focus (DOF) poses significance on imaging subcellular structure of human internal organs. We review several important techniques of DOF extension suitable for miniature probes, among which the probe based on mode interference is believed to have great potential because of its easy fabrication, compact structure, high light transmission efficiency, optimized working distance and DOF, and uniformity of axial light intensity.
    Qiu Jianrong, Han Tao, Wang Di, Meng Jia, Liu Zhiyi, Ding Zhihua. Probes for Endoscopic Optical Coherence Tomography: Minimized Design and Depth of Focus Extension[J]. Chinese Journal of Lasers, 2020, 47(2): 207013
    Download Citation