• Chinese Journal of Quantum Electronics
  • Vol. 23, Issue 4, 522 (2006)
Xing-hua TU1、2、*, Wen-qing LIU1, Tie-dong WANG1, Xiao-mei WANG1, Min WANG1, Yu-jun ZHANG1, Jian-guo LIU1, and Feng-zhong DONG1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: Cite this Article
    TU Xing-hua, LIU Wen-qing, WANG Tie-dong, WANG Xiao-mei, WANG Min, ZHANG Yu-jun, LIU Jian-guo, DONG Feng-zhong. Noise analysis of second-harmonic detection with tunable diode laser absorption spectroscopy[J]. Chinese Journal of Quantum Electronics, 2006, 23(4): 522 Copy Citation Text show less

    Abstract

    As a new advanced method of measuring trace gases,the technology of tunable diode laser absorption spectroscopy(TDLAS)has been applied in the fields of atmospheric chemistry and pollutant gases monitoring. When the absorption optic length can not be effective increased in the system,the noise of system will limit the capabilities of the system and should be reduced to increase the detection limit of TDLAS system. The origin of noise of TDLAS system is introduced and the second-harmonic detection of CO and CO2 with shortpath optic length at atmosphere pressure in the near-infrared wavelength also presented. The detection limit of CO is 0.73% and the one of CO2 is 0.98%. It is obvious that the TDLAS system is fit for the environmental monitor with the high detection limit.
    TU Xing-hua, LIU Wen-qing, WANG Tie-dong, WANG Xiao-mei, WANG Min, ZHANG Yu-jun, LIU Jian-guo, DONG Feng-zhong. Noise analysis of second-harmonic detection with tunable diode laser absorption spectroscopy[J]. Chinese Journal of Quantum Electronics, 2006, 23(4): 522
    Download Citation