    Du Xinghu, Xue Ying, He Xing, Wang Shuai, Yang Ping, Xu Bing. High Reflectivity Measurement Based on Coupled Cavity Ring-Down Technique[J]. Chinese Journal of Lasers, 2020, 47(6): 604006 Copy Citation Text show less
    Schematics of coupled cavity ring-down technique. (a) Initial cavity; (b) measured cavity
    Fig. 1. Schematics of coupled cavity ring-down technique. (a) Initial cavity; (b) measured cavity
    Simulation diagrams of coupled intracavity mode. (a) Optimal coupling state; (b) x-axis offset; (c) y-axis offset; (d) x-axis and y-axis offset to the same degree
    Fig. 2. Simulation diagrams of coupled intracavity mode. (a) Optimal coupling state; (b) x-axis offset; (c) y-axis offset; (d) x-axis and y-axis offset to the same degree
    Schematic of coupled cavity ring-down system for high reflectivity measurement
    Fig. 3. Schematic of coupled cavity ring-down system for high reflectivity measurement
    Experimental diagram of coupled cavity ring-down system for high reflectivity measurement
    Fig. 4. Experimental diagram of coupled cavity ring-down system for high reflectivity measurement
    Facula model of traditional cavity ring-down. (a) Initial cavity being in TEM00 model; (b) measured cavity being in TEM00 model
    Fig. 5. Facula model of traditional cavity ring-down. (a) Initial cavity being in TEM00 model; (b) measured cavity being in TEM00 model
    Facula model of coupled cavity ring-down. (a) Coupled cavity being out of alignment 1; (b) coupled cavity being out of alignment 2; (c) coupled initial cavity being in TEM00 model; (d) coupled measured cavity being in TEM00 model
    Fig. 6. Facula model of coupled cavity ring-down. (a) Coupled cavity being out of alignment 1; (b) coupled cavity being out of alignment 2; (c) coupled initial cavity being in TEM00 model; (d) coupled measured cavity being in TEM00 model
    SampleCavitylength /mCavity loss /10-6Reflectivity /%
    Initial cavityMeasured cavity
    1#0.509947.51523.999.94231±0.00203
    940.91516.799.94237±0.00148
    2#0.9211712.31963.599.97487±0.00310
    1696.61947.399.97488±0.00279▲▲
    Notes:represents the measurement of sample 1# under the coupled cavity ring-down technique;▲▲represents the measurement of sample 2 # under the coupled cavity ring-down technique.
    Table 1. Comparison of measurements obtained by two methods
    Du Xinghu, Xue Ying, He Xing, Wang Shuai, Yang Ping, Xu Bing. High Reflectivity Measurement Based on Coupled Cavity Ring-Down Technique[J]. Chinese Journal of Lasers, 2020, 47(6): 604006