• Acta Optica Sinica
  • Vol. 23, Issue 11, 1379 (2003)
[in Chinese]*, [in Chinese], [in Chinese], and [in Chinese]
Author Affiliations
  • [in Chinese]
  • show less
    DOI: Cite this Article Set citation alerts
    [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Sal inity Measurement Based on Position Sensitive Detector and Optical Technology[J]. Acta Optica Sinica, 2003, 23(11): 1379 Copy Citation Text show less

    Abstract

    The developed salinity measurement setup exploits beam deviation caused by refraction due to the salinity of measured water at the receiving end face of a measurement cell. Based on a differential refractometry, a reference water cell filled with distilled water to reduce influence of temperature drift effectively is partitioned by an oblique optical-transmission plate from a horizontal rectangular water tank, which, as a measurement element, is filled with measured salt water. Beam deviation caused by salinity change is detected by a position sensitive detector. Experimental results indicate the feasibility of the developed system with a measurement resolution of 0.012‰ and repeatability error less than ±0.3‰. Errors which will affect the measurement performance are discussed and analyzed in detail. The results indicate that temperature and light power drifts have little influence on the measurement accuracy.
    [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Sal inity Measurement Based on Position Sensitive Detector and Optical Technology[J]. Acta Optica Sinica, 2003, 23(11): 1379
    Download Citation