• Acta Optica Sinica
  • Vol. 40, Issue 22, 2212005 (2020)
Wenjie Jia, Zhenhua Xi, Dong Fan, Meng Dong, Chengyao Wu, and Yongjun Cheng*
Author Affiliations
  • Research Center of Vacuum Measurement Technology, Lanzhou Institute of Physics, Lanzhou, Gansu 730000, China
  • show less
    DOI: 10.3788/AOS202040.2212005 Cite this Article Set citation alerts
    Wenjie Jia, Zhenhua Xi, Dong Fan, Meng Dong, Chengyao Wu, Yongjun Cheng. Quantum Vacuum Measurement Based on Fabry-Perot Laser Resonant Cavity[J]. Acta Optica Sinica, 2020, 40(22): 2212005 Copy Citation Text show less
    Structural diagram of quantum vacuum measurement device based on Fabry-Perot resonant cavity
    Fig. 1. Structural diagram of quantum vacuum measurement device based on Fabry-Perot resonant cavity
    Test results of laser frequency uncertainty. (a) Variation of beat frequency with time; (b) evaluation result of Allan deviation
    Fig. 2. Test results of laser frequency uncertainty. (a) Variation of beat frequency with time; (b) evaluation result of Allan deviation
    Relationship between measured and retrieved vacuum degrees
    Fig. 4. Relationship between measured and retrieved vacuum degrees
    SerialnumberTime /sΔν /MHzν /MHz
    100384350383.4336
    240000.0007384350383.4329
    380000.0010384350383.4326
    4120000.0017384350383.4319
    5160000.0021384350383.4315
    6200000.0025384350383.4311
    7240000.0030384350383.4306
    8280000.0034384350383.4301
    9320000.0039384350383.4297
    10360000.0044384350383.4292
    Table 1. Partial test result of laser frequency
    Serial numberVacuum degree /Pan-1Serial numberVacuum degree /Pan-1
    11112.73023×10-712100512.51911×10-5
    26201.54968×10-613202305.07167×10-5
    310342.58229×10-614300897.54434×10-5
    422075.52102×10-615402551.00935×10-4
    531697.93029×10-616503041.26143×10-4
    639981.00120×10-517604981.51719×10-4
    750401.26219×10-518700381.75653×10-4
    860671.51961×10-519801922.01135×10-4
    970381.76357×10-520901042.26002×10-4
    1080442.01531×10-5211001862.51302×10-4
    1190462.26681×10-5
    Table 2. Measurement results of refractive index
    Measured value /PaBeat frequency /MHzDeviation of modulusRetrieved value /PaRelative deviation
    10056-595.357100523.90×10-4
    20235227.8613202332.40×10-5
    30094744.0019300849.60×10-5
    402601554.0725402481.21×10-4
    50309757.3332502981.06×10-4
    605031599.6738604967.50×10-5
    700431811.0044700385.10×10-5
    801971119.605180198-7.00×10-6
    901091689.905790113-3.50×10-5
    100191928.3064100199-8.40×10-5
    Table 3. Measured and retrieved results of vacuum degree
    Wenjie Jia, Zhenhua Xi, Dong Fan, Meng Dong, Chengyao Wu, Yongjun Cheng. Quantum Vacuum Measurement Based on Fabry-Perot Laser Resonant Cavity[J]. Acta Optica Sinica, 2020, 40(22): 2212005
    Download Citation