• Laser & Optoelectronics Progress
  • Vol. 51, Issue 3, 31204 (2014)
Zhao Qi1、*, Jiang Zewei1, Meng Qing′an1, Li Yiguo1, Fan Hongying1, Hu Shaoyun1, Li Jianxin2, He Yong2, and Yang Jie3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 10.3788/lop51.031204 Cite this Article Set citation alerts
    Zhao Qi, Jiang Zewei, Meng Qing′an, Li Yiguo, Fan Hongying, Hu Shaoyun, Li Jianxin, He Yong, Yang Jie. Calibration of Foci-Meter Based on Talbot-Moiré Interferometer[J]. Laser & Optoelectronics Progress, 2014, 51(3): 31204 Copy Citation Text show less

    Abstract

    It developed a method to calibrate the foci- meter based on Talbot- Moiré interferometer, the calibration range of focal length up to 100 meters. Moreover, it introduced a foci-meter based on Talbot-Moiré interferometer. It includes a lens group and a laser source to generate laser beams with steady focal length and wavefronts which used to calibrate the foci- meter. The focal length of the lens group is measured by experiments taking advantage of a traceable lens.The following conclusions can be derived from experimental results: the uncertainty of this calibration method is acceptable.
    Zhao Qi, Jiang Zewei, Meng Qing′an, Li Yiguo, Fan Hongying, Hu Shaoyun, Li Jianxin, He Yong, Yang Jie. Calibration of Foci-Meter Based on Talbot-Moiré Interferometer[J]. Laser & Optoelectronics Progress, 2014, 51(3): 31204
    Download Citation