• Acta Optica Sinica
  • Vol. 32, Issue 7, 730002 (2012)
Wei Minxi*, Yang Guohong, Yi Tao, Hou Lifei, Liu Shenye, and Li Jun
Author Affiliations
  • [in Chinese]
  • show less
    DOI: 10.3788/aos201232.0730002 Cite this Article Set citation alerts
    Wei Minxi, Yang Guohong, Yi Tao, Hou Lifei, Liu Shenye, Li Jun. Calibration Research of Integral Diffraction Efficiency of Crystal in Laboratory[J]. Acta Optica Sinica, 2012, 32(7): 730002 Copy Citation Text show less

    Abstract

    Based on stability and angle precision control of the X-ray diffractometer, the calibration technique of integral diffraction efficiency of the flat crystal is introduced by taking the flat PET crystal as sample. X-ray source is Cu X-ray tube and it is monochromatic at Cu-Kα lines by choosing appropriate high voltage and nickel foils to attenuate bremsstrahlung and Cu-Kβ lines. The slit width in front of proportional counter is 0.05 mm. The source and diffractive curve are scanned in steps of 0.001°. The data show that the integral diffraction efficiency is (1.759±0.002)×10-4 rad at Cu-Kα (8047.823 eV). The results show that this calibration technique is efficient and it can be completed conveniently in common laboratory.