• Acta Optica Sinica
  • Vol. 37, Issue 8, 0811003 (2017)
Jingyu Wang, Yu Han, Lei Li, Xiaoqi Xi, Jianbang Liu, and Bin Yan*
Author Affiliations
  • College of Information System Engineering, Information Engineering University, Zhengzhou, Henan 450002, China
  • show less
    DOI: 10.3788/AOS201737.0811003 Cite this Article Set citation alerts
    Jingyu Wang, Yu Han, Lei Li, Xiaoqi Xi, Jianbang Liu, Bin Yan. An Iterative All-Geometric-Parameter Calibration Method for Cone-Beam Computed Laminography System[J]. Acta Optica Sinica, 2017, 37(8): 0811003 Copy Citation Text show less
    Scanning configuration of the cone beam CL system
    Fig. 1. Scanning configuration of the cone beam CL system
    Projections of phantom
    Fig. 2. Projections of phantom
    Convergence curve of F
    Fig. 3. Convergence curve of F
    Final relative error
    Fig. 4. Final relative error
    Reconstruction results of Shepp-Logan phantom. (a) Images of XY slice; (b) images of XZ slice; (c) images of YZ slice
    Fig. 5. Reconstruction results of Shepp-Logan phantom. (a) Images of XY slice; (b) images of XZ slice; (c) images of YZ slice
    Reconstruction results of the PCB phantom. (a) Ideal image; (b) image corrected by Zhang's method; (c) image corrected by the proposed method
    Fig. 6. Reconstruction results of the PCB phantom. (a) Ideal image; (b) image corrected by Zhang's method; (c) image corrected by the proposed method
    ParameterZhang's method in CT systemZhang's method in CL system
    Real valueMeasured valueReal valueMeasured value
    η /(°)0.0350.0340.0290.028
    pX /mm4.5874.59539.86139.657
    pZ /mm0.6640.8671212.9321205.948
    sX /mm55.03954.836
    sY /mm14501450.02514501445.835
    sZ /mm-2-0.456-23.771
    t7.2507.25212.24212.209
    Table 1. Parameter calibration results with noise level of 0.05
    ParameterZhang's method in CT systemZhang's method in CL system
    Real valueMeasured valueReal valueMeasured value
    η /(°)0.0350.0340.0290.028
    pX /mm4.5874.57339.86139.649
    pZ /mm0.6640.3601212.9321206.385
    sX /mm55.04955.529
    sY /mm14501449.78114501447.381
    sZ /mm-2-1.362-25.069
    t7.2507.25212.24212.201
    Table 2. Parameter calibration results with noise level of 0.1
    ParameterReal valueZhang's methodProposed method
    η /(°)0.0290.0280.028
    pX /mm39.86139.65739.742
    pZ /mm1212.9321205.9481212.919
    sX /mm54.8365.055
    sY /mm14501445.8351451.483
    sZ /mm-23.771-0.261
    t12.24212.20912.231
    Table 3. Parameter calibration results with noise level of 0.05
    ParameterReal valueZhang's methodProposed method
    η /(°)0.0290.0280.028
    pX /mm39.86139.64939.673
    pZ /mm1212.9321206.3851212.971
    sX /mm55.5295.089
    sY /mm14501447.3811450.921
    sZ /mm-25.069-0.929
    t12.24212.20112.225
    Table 4. Parameter calibration results with noise level of 0.1
    Jingyu Wang, Yu Han, Lei Li, Xiaoqi Xi, Jianbang Liu, Bin Yan. An Iterative All-Geometric-Parameter Calibration Method for Cone-Beam Computed Laminography System[J]. Acta Optica Sinica, 2017, 37(8): 0811003