• Acta Optica Sinica
  • Vol. 39, Issue 2, 0228002 (2019)
Shun Wang1、*, Hao Zhang1, Wei Li2, Shixin Li1, Jianbo Zhao3, Qingqian Guo1, Ling Wang1, Junjuan Shang1, Xin Li2, and Jiandong Hu1、*
Author Affiliations
  • 1 College of Mechanical & Electrical Engineering, Henan Agricultural University, Zhengzhou, Henan 450002, China;
  • 2 College of Sciences, Henan Agricultural University, Zhengzhou, Henan 450002, China
  • 3 Department of Material and Chemical Engineering, Zhengzhou University of Light Industry, Zhengzhou, Henan 450001, China
  • show less
    DOI: 10.3788/AOS201939.0228002 Cite this Article Set citation alerts
    Shun Wang, Hao Zhang, Wei Li, Shixin Li, Jianbo Zhao, Qingqian Guo, Ling Wang, Junjuan Shang, Xin Li, Jiandong Hu. Construction and Experimental Study of a Multi-Channel Localized Surface Plasmon Resonance Analysis Device[J]. Acta Optica Sinica, 2019, 39(2): 0228002 Copy Citation Text show less
    (a) Schematic of multi-channel precision alignment LSPR experimental setup; (b) multi-channel precision alignment system
    Fig. 1. (a) Schematic of multi-channel precision alignment LSPR experimental setup; (b) multi-channel precision alignment system
    Absorbance distribution of AuNPs with diameter of 13.5 nm at the corresponding positions of microplates
    Fig. 2. Absorbance distribution of AuNPs with diameter of 13.5 nm at the corresponding positions of microplates
    Extinction spectra of AuNPs with different diameters and the relationship between LSPR peak wavelength and the diameter of AuNPs in water medium. (a) Extinction spectra; (b) relationship
    Fig. 3. Extinction spectra of AuNPs with different diameters and the relationship between LSPR peak wavelength and the diameter of AuNPs in water medium. (a) Extinction spectra; (b) relationship
    Extinction spectra of AuNPs with different diameters and the relationship between LSPR peak wavelength and the diameter of AuNPs in glucose solution medium with mass fraction of 5.6%. (a) Extinction spectra; (b) relationship
    Fig. 4. Extinction spectra of AuNPs with different diameters and the relationship between LSPR peak wavelength and the diameter of AuNPs in glucose solution medium with mass fraction of 5.6%. (a) Extinction spectra; (b) relationship
    Relation between AuNPs LSPR peak wavelength and environmental refractive index
    Fig. 5. Relation between AuNPs LSPR peak wavelength and environmental refractive index
    Amount of 1% sodium citrate /mLAmount of 1 mmol/L HAuCl4/mLMolar ratio of sodium citrate to HAuCl4Diameter of prepared AuNPs /nm
    2.50253.87∶113.5
    1.28251.98∶125.5
    0.85251.32∶141.0
    Table 1. Reagents required for AuNPs synthesis and the diameter of AuNPs
    D /nmp1p2p3p4p5Adjusted R2
    5.03.320×10-8-6.920×10-50.05464-18.9724630.9997
    13.56.758×10-8-14.140×10-50.11050-38.2849560.9976
    25.54.436×10-8-9.405×10-50.07451-26.1434300.9990
    41.05.004×10-8-10.650×10-50.08466-29.8239250.9979
    Table 2. Fitting coefficient for different sizes of AuNPs in water medium
    D /nmp1p2p3p4p5Adjusted R2
    5.03.324×10-8-6.77×10-50.05314-18.4724010.9997
    13.58.093×10-8-16.94×10-50.13250-45.9359510.9964
    25.54.844×10-8-10.27×10-50.08133-28.5437430.9987
    41.05.396×10-8-11.49×10-50.09138-32.1942400.9985
    Table 3. Fitting coefficient for different sizes of AuNPs in glucose solution medium with mass fraction of 5.6%
    nrp1p2p3p4p5Adjusted R2
    1.33054.436×10-8-9.41×10-50.07451-26.1434300.9990
    1.34905.097×10-8-10.80×10-50.08555-30.0139350.9989
    1.36555.301×10-8-11.24×10-50.08910-31.2841050.9989
    1.37355.924×10-8-12.56×10-50.09957-34.9545860.9986
    Table 4. Fitting coefficient for AuNPs with diameter of 25.5 nm under different refractive index conditions
    nrp1p2p3p4p5Adjusted R2
    1.33055.004×10-8-10.65×10-50.08466-29.8239250.9979
    1.34905.150×10-8-10.97×10-50.08733-30.7940580.9982
    1.36555.657×10-8-12.05×10-50.09599-33.8644650.9980
    1.37355.872×10-8-12.52×10-50.09976-35.2146450.9986
    Table 5. Fitting coefficient for AuNPs with diameter of 41.0 nm under different refractive index conditions
    Shun Wang, Hao Zhang, Wei Li, Shixin Li, Jianbo Zhao, Qingqian Guo, Ling Wang, Junjuan Shang, Xin Li, Jiandong Hu. Construction and Experimental Study of a Multi-Channel Localized Surface Plasmon Resonance Analysis Device[J]. Acta Optica Sinica, 2019, 39(2): 0228002
    Download Citation