    DOI: 10.3788/AOS202040.1014003 Cite this Article Set citation alerts
    Ying Liu, Yaliang Yang, Xian Yue. Laser Exposure Safety Analysis for Adaptive Optics Retinal Imaging System[J]. Acta Optica Sinica, 2020, 40(10): 1014003 Copy Citation Text show less
    Schematic of human retinal AO imaging system
    Fig. 1. Schematic of human retinal AO imaging system
    Distribution of spectral power density. (a) Imaging light source; (b) beacon light source
    Fig. 2. Distribution of spectral power density. (a) Imaging light source; (b) beacon light source
    Wavelength /nmR(λ)Cc
    700--105010(700-λ)/5001.0
    1050--11500.21.0
    1150--12000.2×100.02×(1150-λ)100.018(λ-1150)
    1200--14000.028+100.04(λ-1250)
    Table 1. Retinal hazard spectral weight function and absorption coefficient[13-14]
    Conditionα /radt /sEL /(W·m-2·sr-1)
    αminα<αmaxt<10-6
    Basic exposure limitαminα<αmax10-6t<0.252×104×α-1×t0.75
    αminα<αmaxt≥0.252.8×104×α-1
    ααmaxt<10-6
    Large sourceααmax10-6t<625×10-6
    ααmax625×10-6t<0.25
    ααmaxt≥0.2528×104
    α<αmint<10-6
    Small sourceα<αmin10-6t<0.251.3×107×t0.75
    α<αmint≥0.251.9×107
    Table 2. Calculation of retinal radiance limits[13]
    α /radNumber of pulses nExposure limit /(W·m-2·sr-1)
    ααmaxn<40EL×n-0.25
    n≥400.EL
    αmax<α≤0.1n<625EL×n-0.25
    n≥6250.EL
    α>0.1EL
    Table 3. Retinal radiance limit for repeated pulse exposure