• Spectroscopy and Spectral Analysis
  • Vol. 29, Issue 5, 1371 (2009)
ZHAO Yu-huan1、2、*, YAN Feng1, LOU Hong-wei1, SUI Yong-xin1, YANG Huai-jiang1, and CAO Jian-lin1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: Cite this Article
    ZHAO Yu-huan, YAN Feng, LOU Hong-wei, SUI Yong-xin, YANG Huai-jiang, CAO Jian-lin. Measurement Technology for Relative Spectral Responsivity of the Ultraviolet ICCD[J]. Spectroscopy and Spectral Analysis, 2009, 29(5): 1371 Copy Citation Text show less

    Abstract

    The spectral responsivity is one of the important parameters of detector. With the development of ultraviolet detect technique, accurate measure for the spectral responsivity of ultraviolet detector becomes more and more important. A measure principle which is based on the direct comparison measuring method for relative spectral response measurement of the ultraviolet ICCD is introduced. Based on the scientific-grade spectrometer with ultraviolet responsivity, a measurement facility was built. The relative spectral responsivity curve of the ultraviolet ICCD was obtained. The curve indicates that the spectral response of UV-ICCD is from 220 to 300 nm and the peak response is near 270 nm, which indicates that the UV-ICCD is solar blind. The analysis of the uncertainty shows that the most largest uncertainty of measurement for relative spectral responsivity of the UV-ICCD is 7.79% and meets the requirement.
    ZHAO Yu-huan, YAN Feng, LOU Hong-wei, SUI Yong-xin, YANG Huai-jiang, CAO Jian-lin. Measurement Technology for Relative Spectral Responsivity of the Ultraviolet ICCD[J]. Spectroscopy and Spectral Analysis, 2009, 29(5): 1371
    Download Citation