• Optics and Precision Engineering
  • Vol. 19, Issue 12, 2870 (2011)
QI Xiang-dong1,*, HAN Peng-peng1, PAN Ming-zhong1, and CUI Ji-cheng1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/ope.20111912.2870 Cite this Article
    QI Xiang-dong, HAN Peng-peng, PAN Ming-zhong, CUI Ji-cheng. Spectral calibration of imaging spectrometer with convex grating[J]. Optics and Precision Engineering, 2011, 19(12): 2870 Copy Citation Text show less

    Abstract

    A spectral calibration facility consisting of a monochromator and a collimator was established to calibrate the Offner imaging spectrometer with a convex grating designed by ourselves. The self-designed software for sampling and processing spectral data was used to calibrate the Offner imaging spectrometer and analyze obtained data. Calibrated results show that the calibration facility has a compact conformation, better commonality and high calibration precision, and its spectral range is better than 400-800 nm and spectral resolution is superior to 10 nm, which can satisfy the calibration need in practice.
    QI Xiang-dong, HAN Peng-peng, PAN Ming-zhong, CUI Ji-cheng. Spectral calibration of imaging spectrometer with convex grating[J]. Optics and Precision Engineering, 2011, 19(12): 2870
    Download Citation