• Chinese Journal of Lasers
  • Vol. 47, Issue 3, 311001 (2020)
Dong Hewei, Guo Ruimin*, Cui Wenchao, and Li Dong
Author Affiliations
  • Environmental Measurement Center, National Institute of Metrology, Beijing 100029, China
  • show less
    DOI: 10.3788/CJL202047.0311001 Cite this Article Set citation alerts
    Dong Hewei, Guo Ruimin, Cui Wenchao, Li Dong. Cavity Ring-Down Spectroscopy Based on Folded Cavity[J]. Chinese Journal of Lasers, 2020, 47(3): 311001 Copy Citation Text show less

    Abstract

    In this paper, to reduce the cavity volume of a cavity ring-down spectrometer (CRDS), we propose designing a Z-folding cavity that has an external dimension of 26.4 cm×8.5 cm×4.5 cm and an expanded length of up to 73.8 cm. In the study, we perform a simulation based on the parallel-Gaussian-beam-transfer-equation to match the laser mode and cavity mode. By adjusting the focal length and position of two convex lenses, we realize mode matching and identify that only the TEM00 mode existes in the cavity. In the experiment, an acousto-optic-modulator is triggered by a digital-delay-generator to switch the laser beam. The feasibility of the Z-folding cavity is verified by the successfully acquiring ring-down signals. By exponentially fitting the experimental data, the ring-down time is determined to be 0.852 μs, which provides a maximum residual of 0.004 μW. The fitted ring-down time is consistent with the theoretical calculation result. The Z-folding cavity designed in this paper is compact and can be used in a commercial CRDS setup.
    Dong Hewei, Guo Ruimin, Cui Wenchao, Li Dong. Cavity Ring-Down Spectroscopy Based on Folded Cavity[J]. Chinese Journal of Lasers, 2020, 47(3): 311001