• Acta Optica Sinica
  • Vol. 28, Issue 12, 2325 (2008)
Li Jianjun*, Zheng Xiaobing, Lu Yunjun, Zhang Wei, and Zou Peng
Author Affiliations
  • [in Chinese]
  • show less
    DOI: Cite this Article Set citation alerts
    Li Jianjun, Zheng Xiaobing, Lu Yunjun, Zhang Wei, Zou Peng. Optical Radiometric Calibration Based on Parametric Down-Conversion Effect[J]. Acta Optica Sinica, 2008, 28(12): 2325 Copy Citation Text show less

    Abstract

    A new calibration method of detectors by using correlated photons generated in spontaneous parametric down-conversion effect of nonlinear crystal is proposed. A CW ultraviolet (351 nm), diode-pumped, frequency-doubled, solid-state laser is used to pump BBO crystal. The degenerate parametric down-conversion (PDC) photons with wavelengths of 702 nm were produced and by tuning crystal angle, nondegenerate PDC photons with wavelengths of 633nm and 789nm were also produced. The quantum efficiencies of the photomultiplier at the above wavelengths were measured respectively. An absolute calibration system of detector quantum efficiency was performed. And its principle and experimental setup were introduced. The calibration uncertainties of PDC method were analyzed systematically. The results show the total calibration relative uncertainties are less than 5.5%. In the end, a comparison between PDC effect method and conventional method based on standard detectors was performed. The results show that the relative difference of the two methods is less than 10%.
    Li Jianjun, Zheng Xiaobing, Lu Yunjun, Zhang Wei, Zou Peng. Optical Radiometric Calibration Based on Parametric Down-Conversion Effect[J]. Acta Optica Sinica, 2008, 28(12): 2325
    Download Citation