    On-board calibration process on VIMI
    Fig. 1. On-board calibration process on VIMI
    Working schematic diagram of SDRDM
    Fig. 2. Working schematic diagram of SDRDM
    Change curve of SDRDM solar observation port signal with azimuth
    Fig. 3. Change curve of SDRDM solar observation port signal with azimuth
    H factor change curve with time and azimuth
    Fig. 4. H factor change curve with time and azimuth
    H factor curve after correction for B3 and B5 bands
    Fig. 5. H factor curve after correction for B3 and B5 bands
    Date|Azimuth /(°)B1B3BrB5
    2018/8/28|157.7391111
    2019/4/16|158.9670.99610.98990.99170.9905
    2019/5/16|161.2061.00760.99260.99320.9926
    2019/8/28|159.8461.00730.99841.00201.0025
    2020/3/17|159.4151.00461.00161.00501.0076
    Average0.00490.00500.00580.0071
    Table 1. H factor under the angle of similar azimuth
    SourceUncertainty /%
    Relative SD BRDF measurement (μ1)0.50
    Relative SD cosine error of incidence (μ2)0.54
    Relationship between transmittance distribution and incident angle (μ3)0.35
    Stability of SDRDM (μ4)0.10
    Stray light (μ5)0.27
    Total uncertainty (U)0.86
    Table 2. Uncertainty of SDRDM monitoring measurement
    SourceUncertainty /%
    SD BRDF measurement (μ1)1
    SD cosine error of incidence (μ2)0.54
    SD radiance non-uniformity (μ3)0.90
    Earth-sun distance correction factor (μ4)0.20
    SD stability monitor (μ5)0.86
    Measurement of SD H factor before launch (μ6)0.50
    Solar spectral irradiance (μ7)1
    Stray light (μ8)2.72
    Total uncertainty (U)3.40
    Table 3. Uncertainty of the SD on-board emitted spectrum radiance