• Acta Optica Sinica
  • Vol. 37, Issue 12, 1234001 (2017)
Li Gu1、2, Fangke Zong1、2, Xiang Li3, Junlan Zhou1, Qinlao Yang1、*, and Baoping Guo1
Author Affiliations
  • 1 Key Laboratory of Optoelectronic Devices and Systems of Ministry of Education and Guangdong Province, College of Optoelectronic Engineering, Shenzhen University, Shenzhen, Guangdong 518060, China
  • 2 College of Information Engineering, Shenzhen University, Shenzhen, Guangdong 518060, China
  • 3 Institute for Advanced Study, Shenzhen University, Shenzhen, Guangdong 518060, China
  • show less
    DOI: 10.3788/AOS201737.1234001 Cite this Article Set citation alerts
    Li Gu, Fangke Zong, Xiang Li, Junlan Zhou, Qinlao Yang, Baoping Guo. Development of X-Ray Streak Camera with Large Dynamic Range and High Temporal-Spatial Resolution[J]. Acta Optica Sinica, 2017, 37(12): 1234001 Copy Citation Text show less
    Electrode design of streak image converter tube
    Fig. 1. Electrode design of streak image converter tube
    Axis potential distribution of streak image converter tube
    Fig. 2. Axis potential distribution of streak image converter tube
    Newly developed streak image converter tube
    Fig. 3. Newly developed streak image converter tube
    Schematic of static testing for streak image converter tube
    Fig. 4. Schematic of static testing for streak image converter tube
    Static spatial resolution image of streak image converter tube
    Fig. 5. Static spatial resolution image of streak image converter tube
    (a) Schematic and (b) photo of dynamic testing for streak camera
    Fig. 6. (a) Schematic and (b) photo of dynamic testing for streak camera
    Dynamic spatial resolution images of streak camera
    Fig. 7. Dynamic spatial resolution images of streak camera
    Dynamic temporal resolution images of streak camera
    Fig. 8. Dynamic temporal resolution images of streak camera
    Sweep speed images of streak camera. (a) S1; (b) S2; (c) S3; (d) S4
    Fig. 9. Sweep speed images of streak camera. (a) S1; (b) S2; (c) S3; (d) S4
    Dynamic range image of streak camera
    Fig. 10. Dynamic range image of streak camera
    Electrode voltage /kVLength /mmMagnificationDeflection sensitivity /(mm·kV-1)
    P/CMF1A1F2A2P/S
    010.426.1510.421.8912124841.6060.71
    Table 1. Voltage and structure parameters of streak image converter tube
    Resolution /(lp·mm-1)Distance /mmCTF
    221.20.206
    184.70.305
    128.20.653
    1011.70.753
    Table 2. Static spatial resolution data of streak image converter tube
    Resolution /(lp·mm-1)Distance /mmCTF
    201.20.162
    164.70.253
    128.20.536
    1011.70.620
    Table 3. Dynamic spatial resolution data of streak camera
    Optical path difference /mmTime interval /psInterval /pixelWidth 1 /pixelWidth 2 /pixelAverage width /pixelTime resolution /ps
    361201255.505.005.255.04
    Table 4. Dynamic temporal resolution data of streak camera
    RangeEtalon distance /mm(ps)Average interval /pixelStandard deviation /pixelSweep speed nonlinearity /%Sweep speed /(mm·ns-1)
    S135(233)252.35.82.311.2
    S275(500)262.74.71.85.5
    S3150(1000)257.76.82.62.7
    S4150(1000)140.92.51.81.5
    Table 5. Sweep speed data of streak camera
    Li Gu, Fangke Zong, Xiang Li, Junlan Zhou, Qinlao Yang, Baoping Guo. Development of X-Ray Streak Camera with Large Dynamic Range and High Temporal-Spatial Resolution[J]. Acta Optica Sinica, 2017, 37(12): 1234001
    Download Citation