• Acta Optica Sinica
  • Vol. 30, Issue 1, 310 (2010)
Li Yude1、2、*, Lin Xiaoyan1、2, Pan Qiuli1、2, and Tan Guotai1、2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/aos20103001.0310 Cite this Article Set citation alerts
    Li Yude, Lin Xiaoyan, Pan Qiuli, Tan Guotai. Study on the Transmission Characteristics of a Tapered X-Ray Capillary[J]. Acta Optica Sinica, 2010, 30(1): 310 Copy Citation Text show less

    Abstract

    Based on the ray-tracing theory,a calculation model for X-ray transmission in a tapered X-ray capillary was set up. Using the calculation model,systematic theoretical studies on the transmission characteristics of a tapered capillary were performed. Calculating the intensity distribution,the conclusion was got that the size of X-ray source and the entrance diameter of the tapered capillary affected the form of intensity distribution. We got the change rules of power density gain K with R0 and f2. Taking the equivalent distance Leq as a standard,the optimal design of tapered capillary was discussed.
    Li Yude, Lin Xiaoyan, Pan Qiuli, Tan Guotai. Study on the Transmission Characteristics of a Tapered X-Ray Capillary[J]. Acta Optica Sinica, 2010, 30(1): 310
    Download Citation