Development of a Mid-Infrared Methane and Carbon Dioxide Dual-Gas Sensor System
Qingchuan Zhao
*
DOI:
10.3788/AOS202040.2330001
Cite this Article
Set citation alerts
Qingchuan Zhao. Development of a Mid-Infrared Methane and Carbon Dioxide Dual-Gas Sensor System[J]. Acta Optica Sinica, 2020, 40(23): 2330001
Copy Citation Text
EndNote(RIS)
BibTex
Plain Text
show less
Fig. 1.
Schematic of mid-infrared methane and carbon dioxide sensing system
Download full size
Fig. 2.
Relation between spectra of Lms34 LED, Lms36 PD, Lms43 LED and Lms43 PD and main absorption peaks of methane and carbon dioxide
Download full size
Fig. 3.
Influences of temperature on emission intensity and peak wavelength of Lms34 LED
Download full size
Fig. 4.
LED array design drawing with thermoelectric cooling device
Download full size
Fig. 5.
Emission spectra of Lms34 LED at different pulse currents
Download full size
Fig. 6.
Pulse current driving timing of Lms34 LED and Lms43 LED
Download full size
Fig. 7.
Temperature influence factor curves
Download full size
Fig. 8.
Sensor system experiment platform
Download full size
Temperature /℃
Measurement
value /%
Methane (2.01%)
CO
2
(2.49%)
-20
2.19
2.79
-10
2.14
2.70
0
2.10
2.63
10
2.06
2.57
20
2.01
2.49
30
1.97
2.42
40
1.91
2.36
50
1.87
2.28
Table 1.
System measurement value under standard gas concentration in the temperature range from -20 ℃ to 50 ℃
Parameter
Value
Working voltage /V
3.2-4.0
Average current /mA
<13
CH
4
measuring range
(volume fraction) /%
0-100
CO
2
measuring range
(volume fraction) /%
0-5.00
CH
4
indication error
±0.06%(0-1.00%)
±6% of true concentration
(1.00%-100%)
CO
2
indication error
±0.05%(0-0.50%)
±5%FS (>0.50%-5.00%)
Response time /s
<20
Table 2.
Basic performance index of sensor system
Abstract
Get PDF(in Chinese)
Figures&Tables (10)
Equations (0)
References (20)
Cited By (4)
Get Citation
Copy Citation Text
Qingchuan Zhao. Development of a Mid-Infrared Methane and Carbon Dioxide Dual-Gas Sensor System[J]. Acta Optica Sinica, 2020, 40(23): 2330001
Download Citation
EndNote(RIS)
BibTex
Plain Text
Set citation alerts for the article
Tools
Share
Set citation alerts for the article
Save the article for my favorites
Paper Information
Category: Spectroscopy
Received: Jul. 13, 2020
Accepted: Aug. 26, 2020
Published Online: Nov. 23, 2020
The Author Email: Zhao Qingchuan (zhaoqich@163.com)
DOI:
10.3788/AOS202040.2330001
Recommended Topics
laser devices and laser physics
Lasers and Laser Optics
Laser physics
laser manufacturing
Instrumentation, Measurement and Metrology
Set citation alerts for the article
Please enter your email address
Cancel
Confirm