• Acta Optica Sinica
  • Vol. 38, Issue 11, 1130002 (2018)
Yue Wu1、2、3、*, Jiaxiang Liu1、3, Yonghua Fang1、3、*, and Min Yang1、2、3
Author Affiliations
  • 1 Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Heifei, Anhui 230026, China
  • 3 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    DOI: 10.3788/AOS201838.1130002 Cite this Article Set citation alerts
    Yue Wu, Jiaxiang Liu, Yonghua Fang, Min Yang. Wavelength Calibration of Infrared Spectrometer Based on Hollow Planar Waveguide[J]. Acta Optica Sinica, 2018, 38(11): 1130002 Copy Citation Text show less
    Structural schematics of the hollow planar waveguide infrared spectrometer. (a) Schematic of beam propagation in space; (b) scheme of optical path design
    Fig. 1. Structural schematics of the hollow planar waveguide infrared spectrometer. (a) Schematic of beam propagation in space; (b) scheme of optical path design
    Flow chart of laboratory wavelength calibration
    Fig. 2. Flow chart of laboratory wavelength calibration
    Sketch map of laboratory wavelength calibration
    Fig. 3. Sketch map of laboratory wavelength calibration
    Physical map of wavelength calibration device in laboratory
    Fig. 4. Physical map of wavelength calibration device in laboratory
    Relation of pixel and response at different wavelengths
    Fig. 5. Relation of pixel and response at different wavelengths
    Fitting curves of wavelength calibration
    Fig. 6. Fitting curves of wavelength calibration
    Fitting residual of wavelength calibration
    Fig. 7. Fitting residual of wavelength calibration
    ParameterValue
    Wavelength range /μm8-12
    Spectral sampling interval /μm0.0625
    Numerical aperture0.22
    Detector array size64×1
    Detector pixel size /(mm×mm)0.5×1.5
    Table 1. Specific parameters of spectrometer
    ParameterValue
    Wavelength range /μm9.1-10.9
    Mode>90% TEM00, M2<1.25
    Amplitude stability±1% short term (unstabilized)
    Frequency stability /(kHz·s-1)200
    Beam diameter (1/e2)4.8 mm@output coupler
    Beam divergence /mrad6.5
    Table 2. Parameters of PL2-M CO2 laser
    Central wavelength /μmPixelSpectral resolution /nm
    9.26148.925139
    9.48845.925144
    9.62144.822137
    10.23337.558134
    10.47634.535129
    10.76430.828135
    Table 3. Results of wavelength calibration
    Central Wavelength /μmPixel1st order polynomial fitting3rd order polynomial fitting
    Fitted value /μmError /μmFitted value /μmError /μm
    8.5060.558.284-0.2168.461-0.039
    11.5213.6012.2280.70711.415-0.100
    Table 4. Calculation results of two fitting methods
    Yue Wu, Jiaxiang Liu, Yonghua Fang, Min Yang. Wavelength Calibration of Infrared Spectrometer Based on Hollow Planar Waveguide[J]. Acta Optica Sinica, 2018, 38(11): 1130002
    Download Citation