• Spectroscopy and Spectral Analysis
  • Vol. 34, Issue 1, 274 (2014)
JIANG Cheng-zhi1、2、*, LIANG Jing-qiu3, LIANG Zhong-zhu3, SUN Qiang2, and WANG Wei-biao3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 10.3964/j.issn.1000-0593(2014)01-0274-05 Cite this Article
    JIANG Cheng-zhi, LIANG Jing-qiu, LIANG Zhong-zhu, SUN Qiang, WANG Wei-biao. Influence of Collimation System on Static Fourier Transform Spectrometer[J]. Spectroscopy and Spectral Analysis, 2014, 34(1): 274 Copy Citation Text show less

    Abstract

    Collimation system provides collimated light for the static Fourier-transform spectroscopy (SFTS). Its quality is crucial to the signal to noise ratio (SNR) of SFTS. In the present paper, the physical model of SFTS was established based on the Fresnel diffraction theory by means of numerical software. The influence of collimation system on the SFTS was discussed in detail focusing on the aberrations of collimation lens and the quality of extended source. The results of simulation show that the influences of different kinds of aberrations on SNR take on obvious regularity, and in particular, the influences of off-axis aberrations on SNR are closely related to the location of off-axis point source. Finally the extended sources maximum radius allowed was obtained by simulation, which equals to 0.65 mm. The discussion results will be used for the design of collimation system.
    JIANG Cheng-zhi, LIANG Jing-qiu, LIANG Zhong-zhu, SUN Qiang, WANG Wei-biao. Influence of Collimation System on Static Fourier Transform Spectrometer[J]. Spectroscopy and Spectral Analysis, 2014, 34(1): 274
    Download Citation