• Chinese Journal of Quantum Electronics
  • Vol. 29, Issue 1, 8 (2012)
Hu LI1、2, Wen-qi NIU3, Hong-mei WANG1, Chao-qun HUANG1, Jian-quan LI1, Cheng-yin SHEN1, Hai-he JIANG1, and Yan-nan CHU1、*
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 10.3969/j.issn.1007-5461. 2012.01.002 Cite this Article
    LI Hu, NIU Wen-qi, WANG Hong-mei, HUANG Chao-qun, LI Jian-quan, SHEN Cheng-yin, JIANG Hai-he, CHU Yan-nan. Parameters and performance analysis of photoionization ion mobility spectrometry[J]. Chinese Journal of Quantum Electronics, 2012, 29(1): 8 Copy Citation Text show less

    Abstract

    The construction and performance study is reported for a newly developed photoionization ion mobility spectrometer (UV-IMS) equipped with a 10.6 eV Kr lamp. The effect of the total voltage of drift tube, voltage and pulse width of ion gate on spectral resolution and sensitivity of UV-IMS were discussed. Their optimal values of pulse width, voltage of ion gate and drift tube are 0.15 ms, 80 V, and 2153 V respectively. The concentrations of benzene were prepared by means of exponential dilution method. The experiments show that the UV-IMS has a limit of detection at 400 ppt for benzene and its measurable linear dynamics range is over three orders of magnitude.
    LI Hu, NIU Wen-qi, WANG Hong-mei, HUANG Chao-qun, LI Jian-quan, SHEN Cheng-yin, JIANG Hai-he, CHU Yan-nan. Parameters and performance analysis of photoionization ion mobility spectrometry[J]. Chinese Journal of Quantum Electronics, 2012, 29(1): 8
    Download Citation