• Acta Optica Sinica
  • Vol. 30, Issue 12, 3655 (2010)
Pei Shixin1、*, Cui Fenping1, Zhang Chengyi1, and Li Chuanqi1、2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/aos20103012.3655 Cite this Article Set citation alerts
    Pei Shixin, Cui Fenping, Zhang Chengyi, Li Chuanqi. Detection of Water Vapour Based on Integrated Cavity Output Spectroscopy[J]. Acta Optica Sinica, 2010, 30(12): 3655 Copy Citation Text show less

    Abstract

    A system of integrated cavity output spectroscopy (ICOS) based on a tunable distributed feedback (DFB) diode laser (TDL) is described. In this system a tunable DFB diode laser with the central output wavelength 1315 nm is used as the light source, the length of the absorption cell is 38 cm which is consisted by two mirrors with high reflectivity, near the laser central wavelength 1315 nm, the mirror′s reflectivity is about 99.7%. Presentation of experimental principle, experimental scheme and diode laser wavelength calibration are given; As experimental result, the absorption spectrum of water vapour at 7612.026, 7612.269 and 7610.224 cm-1 are obtained and a detection sensitivity of 2.02×10-7 cm-1 has been achieved. It is experimentally demonstrated that ICOS is a high sensitive and high resolution spectral technology with the advantage of simple experimental setup and easy operation.
    Pei Shixin, Cui Fenping, Zhang Chengyi, Li Chuanqi. Detection of Water Vapour Based on Integrated Cavity Output Spectroscopy[J]. Acta Optica Sinica, 2010, 30(12): 3655
    Download Citation