    Li Xiangxian, Luo Guishan, Xu Liang, Gao Minguang, Tong Jingjing, Wei Xiuli, Liu Na. Standard scale calibration methods for δ13CO2 value measurements based on Fourier Transform Infrared spectroscopy[J]. Infrared and Laser Engineering, 2015, 44(10): 2959 Copy Citation Text show less

    Abstract

    Calibration to standard reference gas scale is necessary to improve the accuracy of the measurements,and ensure the compatibility of measurement values between different instruments for δ13CO2 value based on Fourier Transform Infrared(FTIR) spectroscopy. Firstly, the calibration methods were put forward, the absolute calibration and empirical calibration methods were studied in detail. On the basis of these researches, the δ13CO2 values were calibrated to standard reference gas scale. The calibration results show that measurements accuracy has been improved greatly, compared with measurement results of no calibration. The accuracy for absolute calibration and empirical calibration methods are improved 5.5 and 6.4 times respectively. The study is benefit to improve the accuracy of FTIR measurements and the compatibility between different instruments.
    Li Xiangxian, Luo Guishan, Xu Liang, Gao Minguang, Tong Jingjing, Wei Xiuli, Liu Na. Standard scale calibration methods for δ13CO2 value measurements based on Fourier Transform Infrared spectroscopy[J]. Infrared and Laser Engineering, 2015, 44(10): 2959