• Laser & Optoelectronics Progress
  • Vol. 57, Issue 18, 181701 (2020)
Jinxuan Huang*, Yanfang He, and Feng Lin**
Author Affiliations
  • 2Key Laboratory of Medical Optoelectronic Science and Technology, Ministry of Education, College of Optoelectronics and Information Engineering, Fujian Normal University, Fuzhou, Fujian 350007, China
  • show less
    DOI: 10.3788/LOP57.181701 Cite this Article Set citation alerts
    Jinxuan Huang, Yanfang He, Feng Lin. Dual-Field Capsule Endoscope Design Based on Zemax Software[J]. Laser & Optoelectronics Progress, 2020, 57(18): 181701 Copy Citation Text show less
    Schematics of capsule endoscopes. (a) Traditional capsule endoscope; (b) dual view capsule endoscope
    Fig. 1. Schematics of capsule endoscopes. (a) Traditional capsule endoscope; (b) dual view capsule endoscope
    Initial structure diagram of proposed system
    Fig. 2. Initial structure diagram of proposed system
    Back view structure diagrams. (a) Before reflection; (b) after reflection
    Fig. 3. Back view structure diagrams. (a) Before reflection; (b) after reflection
    Diagram of receiver
    Fig. 4. Diagram of receiver
    Reflection surface diagram
    Fig. 5. Reflection surface diagram
    Diagram of dual view structure
    Fig. 6. Diagram of dual view structure
    Structure diagrams of views. (a) Front view; (b) back view
    Fig. 7. Structure diagrams of views. (a) Front view; (b) back view
    Modulation transfer function of views. (a) Front view; (b) back View
    Fig. 8. Modulation transfer function of views. (a) Front view; (b) back View
    Field curvature and distortion. (a) Front view; (b) back view
    Fig. 9. Field curvature and distortion. (a) Front view; (b) back view
    Modulation transfer function for defocusing. (a) Front view; (b) back view
    Fig. 10. Modulation transfer function for defocusing. (a) Front view; (b) back view
    Optical size /inch (1 inch=2.54 cm)ResolutionPixel size /(μm×μm)Effective area /(mm×mm)
    1/61920×10801.4×1.42.7×1.5
    Table 1. Parameters of imaging chip
    ParameterBack viewconfigurationFront viewconfiguration
    F-number66
    Image height /mm0 to 11 to 1.5
    Working distance /mm1015
    Total length /mm<20<10
    Field of view /(°)8050 to 80
    Table 2. System design parameters
    Jinxuan Huang, Yanfang He, Feng Lin. Dual-Field Capsule Endoscope Design Based on Zemax Software[J]. Laser & Optoelectronics Progress, 2020, 57(18): 181701