  • Vol. 37, Issue 9, 0904001 (2017)
Hui Liu1、2、*, Hongyao Chen1, Xiaolong Si1, and Liming Zhang1
Author Affiliations
  • 1 Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Anhui 230026, China
  • show less
    DOI: 10.3788/AOS201737.0904001 Cite this Article Set citation alerts
    Hui Liu, Hongyao Chen, Xiaolong Si, Liming Zhang. Linear Calibration of Wide Dynamic Range Response of Silicon Detectors Based on Double LEDs Light Source Integrating Sphere[J]. Acta Optica Sinica, 2017, 37(9): 0904001 Copy Citation Text show less
    Schematic of the Si detectors linearity measurement instrument
    Fig. 1. Schematic of the Si detectors linearity measurement instrument
    Relationship between detector response current and LED injection current
    Fig. 2. Relationship between detector response current and LED injection current
    Nonlinearity correction factor of Si detector
    Fig. 3. Nonlinearity correction factor of Si detector
    Linearity error of Si detector
    Fig. 4. Linearity error of Si detector
    Relative spectral radiance of two LEDs
    Fig. 5. Relative spectral radiance of two LEDs
    Relative spectral radiance of the LED with different injection currents
    Fig. 6. Relative spectral radiance of the LED with different injection currents
    Schematic of the measurement of spectral drift
    Fig. 7. Schematic of the measurement of spectral drift
    Timing diagram of LEDs switch and data collection
    Fig. 8. Timing diagram of LEDs switch and data collection
    Uncorrected and corrected nonlinearity factors
    Fig. 9. Uncorrected and corrected nonlinearity factors
    Relative uncertainty of nonlinearity factor
    Fig. 10. Relative uncertainty of nonlinearity factor
    Relative uncertainty of nonlinearity correct factor
    Fig. 11. Relative uncertainty of nonlinearity correct factor
    Injection current0.76~5.3 μA5.3~92 μA92 μA~2.5 mA2.5 mA~0.57 A
    Nonlinearity factor (OD: 3)-1.0000361.0001511.00011
    Nonlinearity factor (OD: 2)1.002210.999781.0001001.00018
    Nonlinearity factor (OD: 1)0.998250.999951.0000481.00018
    Nonlinearity factor (OD: 0.1)1.004800.999781.0000151.00016
    Spectral drift ur,W2.69×10-31.11×10-45.17×10-52.86×10-5
    Table 1. Measurement data of spectral drift
    LED injection currentResponse currentSpectral drifturW /10-5Correction factorurCn /10-5Combined uncertaintyur /10-5
    0.76-5.3 μA39.2 pA-0.326 nA2697.05269
    5.3-92 μA0.326-81.6 nA11.13.2311.6
    92 μA-2.5 mA81.6 nA-20.4 μA5.171.435.36
    2.5 mA-0.57 A20.4 μA-10.2 mA2.861.433.20
    Table 2. Combined uncertainty of measurement system
    Hui Liu, Hongyao Chen, Xiaolong Si, Liming Zhang. Linear Calibration of Wide Dynamic Range Response of Silicon Detectors Based on Double LEDs Light Source Integrating Sphere[J]. Acta Optica Sinica, 2017, 37(9): 0904001