    Manshu Du, Xiaohui Lin, Jiayu Yang, Jianchun Liu, Dongxu Zhang. Calibration Method of Fluorescence Excited Light Path in Nucleic Acid Detection System[J]. Laser & Optoelectronics Progress, 2022, 59(7): 0712003 Copy Citation Text show less
    Structure diagram of optical system of nucleic acid detection instrument
    Fig. 1. Structure diagram of optical system of nucleic acid detection instrument
    Schematic of excitation light path
    Fig. 2. Schematic of excitation light path
    Relation curve between exit angle of light source and radius of light at edge of secondary refraction surface
    Fig. 3. Relation curve between exit angle of light source and radius of light at edge of secondary refraction surface
    Relation curve between exit angle of light source and exit angle of secondary refraction surface
    Fig. 4. Relation curve between exit angle of light source and exit angle of secondary refraction surface
    ZEMAX simulation results
    Fig. 5. ZEMAX simulation results
    Experimental setup for measuring light spot
    Fig. 6. Experimental setup for measuring light spot
    Images taken by CCD with five different apertures
    Fig. 7. Images taken by CCD with five different apertures
    Changes of exit spot radius of excitation light path under different apertures
    Fig. 8. Changes of exit spot radius of excitation light path under different apertures
    Images taken by CCD of optical path collimation experiments under different distances between screen and lens. (a) 7.7 mm; (b) 8.7 mm
    Fig. 9. Images taken by CCD of optical path collimation experiments under different distances between screen and lens. (a) 7.7 mm; (b) 8.7 mm
    r1 /mmr2 /mmn1n2l/mmg/mm
    5.171.01.57.37.7
    Table 1. Parameters table of numerical calculation system for spot radius
    ExperimentDistance /mmSpot radius /dpiReference size /dpiActual size /mmExit angle /(°)
    17.724.9711.42.19-2.86
    8.724.3811.42.14
    27.724.6911.22.20-1.72
    8.724.7911.42.17
    37.724.7911.32.19-0.57
    8.724.6611.32.18
    Table 2. Experimental data of collimation of excitation light path