• Spectroscopy and Spectral Analysis
  • Vol. 42, Issue 4, 1064 (2022)
Zhou-xuan OUYANG*, Ying-jie MA, Dou-dou LI, and Yi LIU
Author Affiliations
  • Chendu University of Technology, Chendu 610059, China
  • show less
    DOI: 10.3964/j.issn.1000-0593(2022)04-1064-06 Cite this Article
    Zhou-xuan OUYANG, Ying-jie MA, Dou-dou LI, Yi LIU. The Research of Polarized Energy Dispersive X-Ray Fluorescence for Measurement Trace Cadmium by Geant4 Simulation[J]. Spectroscopy and Spectral Analysis, 2022, 42(4): 1064 Copy Citation Text show less
    Schematic geometry for polarized beam XRF
    Fig. 1. Schematic geometry for polarized beam XRF
    Comparison of experimental primary spectrum and GEANT4 with 50kV tube voltage
    Fig. 2. Comparison of experimental primary spectrum and GEANT4 with 50kV tube voltage
    Comparison of different anode target angle
    Fig. 3. Comparison of different anode target angle
    Fluorescence detection efficiency of different anode target
    Fig. 4. Fluorescence detection efficiency of different anode target
    Secondary spectrum of Te at 70 kV tube voltage
    Fig. 5. Secondary spectrum of Te at 70 kV tube voltage
    Comparison of primary X-ray active sample and secondary target
    Fig. 6. Comparison of primary X-ray active sample and secondary target
    Spectrum of BaSO4 secondary target
    Fig. 7. Spectrum of BaSO4 secondary target
    Spectra of BaSO4 secondary target at different tube voltage
    Fig. 8. Spectra of BaSO4 secondary target at different tube voltage
    厚度/μm峰面积SNR厚度/μm峰面积SNR
    100.17618.539800.31421.434
    200.23820.0451000.31918.298
    300.27318.7981500.32518.344
    400.28920.5152000.32617.677
    600.30720.5853000.32518.654
    Table 1. SNR of secondary target with different thicknesses at 50 kV tube voltage
    电压/kV峰面积SNR
    500.32617.677
    550.35318.186
    600.37618.884
    650.38920.366
    700.39823.665
    Table 2. SNR of different tube voltages at 200 μm secondary target
    电压/kV峰面积峰背比
    506 43014.179
    557 73015.276
    608 73317.275
    659 43419.512
    709 49421.431
    Table 3. Peak-to-background of different tube voltage
    Zhou-xuan OUYANG, Ying-jie MA, Dou-dou LI, Yi LIU. The Research of Polarized Energy Dispersive X-Ray Fluorescence for Measurement Trace Cadmium by Geant4 Simulation[J]. Spectroscopy and Spectral Analysis, 2022, 42(4): 1064
    Download Citation