• Laser & Optoelectronics Progress
  • Vol. 55, Issue 1, 11201 (2018)
Zheng Xingchun, Zhao Min*, and Feng Shaohua
Author Affiliations
  • School of Mechanical and Precision Instrument Engineering, Xi''an University of Technology,Xi''an, Shaanxi 710048, China
  • show less
    DOI: 10.3788/LOP55.011201 Cite this Article Set citation alerts
    Zheng Xingchun, Zhao Min, Feng Shaohua. Two-Step Calibration of Probe Tip Center of Planar Target[J]. Laser & Optoelectronics Progress, 2018, 55(1): 11201 Copy Citation Text show less
    Vision coordinate measurement system of planar target
    Fig. 1. Vision coordinate measurement system of planar target
    Diagram of target in different positions
    Fig. 2. Diagram of target in different positions
    Simulation results of coordinate T'x and T'y of probe tip center before and after calibration. (a) T'x; (b) T'y
    Fig. 3. Simulation results of coordinate T'x and T'y of probe tip center before and after calibration. (a) T'x; (b) T'y
    Simulation results of corrected value Δx, Δy and Δz of probe tip center
    Fig. 4. Simulation results of corrected value Δx, Δy and Δz of probe tip center
    Simulation results of corrected values Δx and Δy of probe tip center
    Fig. 5. Simulation results of corrected values Δx and Δy of probe tip center
    Relationship between coordinate T″y of probe tip center and angle tilt β of target
    Fig. 6. Relationship between coordinate T″y of probe tip center and angle tilt β of target
    Trend of coordinate T″y of probe tip center
    Fig. 7. Trend of coordinate T″y of probe tip center
    Coordinate location12345678
    x85135175185235285325335
    y00-500050
    z00000000
    Table 1. Coordinates of targets under target coordinate systemmm
    Image numberkProposed methodRef.[8]
    T'xT'yT'zT'xT'yT'z
    1-16.399280.0071931.306-16.495279.8771931.014
    2-16.298280.0401931.864-16.384279.9241931.567
    3-16.300279.9981931.627-16.402279.8891931.327
    4-16.246279.9951931.548-16.368279.8951931.247
    5-16.357279.9921931.567-16.480279.9041931.265
    6-16.429280.0241932.023-16.570279.9501931.717
    7-16.419280.0471931.301-16.577279.9951931.000
    8-16.330280.0781931.769-16.514280.0361931.463
    9-16.303280.0871931.853-16.491280.0571931.545
    10-16.211280.0531932.106-16.407280.0181931.797
    11-16.213280.0771932.156-16.409280.0441931.846
    Mean-16.319280.0361931.738-16.463279.9631931.435
    Standard deviation0.0770.0350.2960.0740.0690.292
    Table 2. Results of probe tip center calibration with the first set of imagesmm
    Zheng Xingchun, Zhao Min, Feng Shaohua. Two-Step Calibration of Probe Tip Center of Planar Target[J]. Laser & Optoelectronics Progress, 2018, 55(1): 11201
    Download Citation