• Acta Optica Sinica
  • Vol. 36, Issue 2, 212005 (2016)
Li Mengfan1、2、3、*, Xu Weiwei1、3, Zou Peng1、3, Zhang Liming1、3, Meng Binghuan1、3, Yang Benyong1、3, Hong Jin1、3, and Qiao Yanli
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 10.3788/aos201636.0212005 Cite this Article Set citation alerts
    Li Mengfan, Xu Weiwei, Zou Peng, Zhang Liming, Meng Binghuan, Yang Benyong, Hong Jin, Qiao Yanli. Response Characteristic of the Ratioing Radiometer[J]. Acta Optica Sinica, 2016, 36(2): 212005 Copy Citation Text show less

    Abstract

    Ratioing radiometer (RR) is a key instrument which can track the on-orbit degradation in bidirectional reflectance distribution function (BRDF) of on-board solar diffuser (SD), and its response to sun angle of incidence (K factor) is an important parameter which need to be measured pre-launch. A brief overview of the principle of RR is provided in this paper, K factor of each band of RR is acquired after several round of tests, which take halogen lamp in the lab instead of sunlight outdoor as light source . Uncertainty of K factors at 470, 650, 825 nm are acquired on the base of the data and error analysis. Result shows that the uncertainty of the K factor at 470 nm is 0.48% and within 0.16% at 650 nm and 825 nm, which meet the requirement of K factor uncertainty, providing valid basis for estimation of on-orbit calibration uncertainty and correction of the calibration coefficient.
    Li Mengfan, Xu Weiwei, Zou Peng, Zhang Liming, Meng Binghuan, Yang Benyong, Hong Jin, Qiao Yanli. Response Characteristic of the Ratioing Radiometer[J]. Acta Optica Sinica, 2016, 36(2): 212005
    Download Citation