• Acta Optica Sinica
  • Vol. 30, Issue s1, 100211 (2010)
Di Qingyue1、2、*, Zhang Wenbo2, Weng Xuan1, and Zhang Xiaoguang1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/aos201030.s100211 Cite this Article Set citation alerts
    Di Qingyue, Zhang Wenbo, Weng Xuan, Zhang Xiaoguang. A New Approach of Calibrating the Polarimeter Module Based on Vector Projection Algorithm[J]. Acta Optica Sinica, 2010, 30(s1): 100211 Copy Citation Text show less

    Abstract

    To solve the problem of calibrating the polarimeter module, a new approach based on vector projection algorithm was proposed, and the principle of calibration was analyzed. The continuous-wave (CW) light passed through the tunable optical attenuator, polarization analyzer, photon detectors of polarimeter module, then the photocurrent was fed into the preamplifier circuit and high-speed analog to digital converter (ADC) in turn. Finally, the digital voltage signals were acquired by the digital signal processor (DSP) and sent to the computer. At the same time, the polarization analyzer exported corresponding Stokes parameters. The calibration matrix could be calculated by the vector projection algorithm based on the given data. The experimental result shows that, in comparison with the common iterative algorithm, the vector projection algorithm has even higher computational efficiency and accuracy. The measured error of the degree of polarization (DOP) is less than 3 percent.
    Di Qingyue, Zhang Wenbo, Weng Xuan, Zhang Xiaoguang. A New Approach of Calibrating the Polarimeter Module Based on Vector Projection Algorithm[J]. Acta Optica Sinica, 2010, 30(s1): 100211