• Optoelectronic Technology
  • Vol. 40, Issue 3, 170 (2020)
Tingting MA1, Jin YANG2, Jiwei ZHU1, Ci SUN2..., Nan SONG2, Yong PENG1, Xueqiang GUO1, Jian ZHANG1, Shulong FENG2 and Hanzhou GUO3|Show fewer author(s)
Author Affiliations
  • 1Changchun SPKTRM Optronics co.Ltd,Changchun 3002,CHN
  • 2Changchun Institute of Optics,Fine Mechanics and Physics,Chinese Academy of Sciences,Changchun 130033,CHN
  • 3Changchun Guoke Medical Engineering Technology Development co.Ltd,Changchun 10000,CHN
  • show less
    DOI: 10.19453/j.cnki.1005-488x.2020.03.004 Cite this Article
    Tingting MA, Jin YANG, Jiwei ZHU, Ci SUN, Nan SONG, Yong PENG, Xueqiang GUO, Jian ZHANG, Shulong FENG, Hanzhou GUO. Study on the Structure of a Miniature Flat⁃field Spectrophotometer Based on Biochemical Analysis[J]. Optoelectronic Technology, 2020, 40(3): 170 Copy Citation Text show less
    Structure of the miniature flat-field spectrophotometer
    Fig. 1. Structure of the miniature flat-field spectrophotometer
    Structural schematic of the flat field concave holographic grating
    Fig. 2. Structural schematic of the flat field concave holographic grating
    Schematic of slot setting error
    Fig. 3. Schematic of slot setting error
    Wavelength error caused by slit setting error
    Fig. 4. Wavelength error caused by slit setting error
    Wavelength error caused by the error of the number of grating lines
    Fig. 5. Wavelength error caused by the error of the number of grating lines
    Schematic of the wavelength error caused by machining error of detector or exit slit
    Fig. 6. Schematic of the wavelength error caused by machining error of detector or exit slit
    Wavelength error caused by machining error of detector or exit slot
    Fig. 7. Wavelength error caused by machining error of detector or exit slot
    Structural schematic of the miniature flat-field spectrophotometer
    Fig. 8. Structural schematic of the miniature flat-field spectrophotometer
    Prototype of the miniature flat-field spectrophotometer
    Fig. 9. Prototype of the miniature flat-field spectrophotometer
    参数名称指标
    刻线密度/(g·mm-1)800
    线分散/(nm·mm-1)11.1
    工作波长范围/nm350~750
    闪耀波长/nm450
    入射角/(°)0
    入射臂/mm86.3
    谱面位置距离/mm85.9
    谱面位置角度/(°)3.54
    Table 1. 平场凹面全息光栅使用参数
    参数名称指标
    光谱带宽/nm8
    工作范围/nm340~800
    波长准确度/nm±2
    通道数16
    Table 2. 小型平像场分光光度计技术指标
    Tingting MA, Jin YANG, Jiwei ZHU, Ci SUN, Nan SONG, Yong PENG, Xueqiang GUO, Jian ZHANG, Shulong FENG, Hanzhou GUO. Study on the Structure of a Miniature Flat⁃field Spectrophotometer Based on Biochemical Analysis[J]. Optoelectronic Technology, 2020, 40(3): 170
    Download Citation