• Spectroscopy and Spectral Analysis
  • Vol. 37, Issue 6, 1778 (2017)
LI Wei1, FAN Xian-guang1、2, WANG Xin1, TANG Ming1, QUE Jing1, HE Jian1, and ZUO Yong3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 103964/jissn1000-0593(2017)06-1778-06 Cite this Article
    LI Wei, FAN Xian-guang, WANG Xin, TANG Ming, QUE Jing, HE Jian, ZUO Yong. Design of Rapid Detection System for Urotropine in Food Based on SERS[J]. Spectroscopy and Spectral Analysis, 2017, 37(6): 1778 Copy Citation Text show less

    Abstract

    Based on surface enhanced Raman spectroscopy (SERS) technology, the detection system for rapid fild determination of urotropine in food is proposed. The detecting system consists of multifunctional pretreating module, Roman optical module and embedded master control module. It completed the pretreatment of sample, the generation, collection and transmission of the Raman spectral data. The multifunctional pretreating module integrate centrifuge, ultrasound and volatile functions all together, reducing the equipments used in the pretreatment experiment. Besides, it is portable and convenient for operation. The Roman optical module we designed embodies the probe of Raman spectrometer and monochromator, and the numerical aperture of the aspherical lens we used in the optical system of the probe was 06. Besides, the monochromator was designed based on Asymmetrical crossed Czerny-Turner structure. The hardware part of the embedded master control module is made up of ARM and FPGA, and its peripheral circuits. The software part of the module is made up of real time operation system μC/OS-Ⅲ, graphics library emWin, file system FatFs and various types of equipment components. Thus, accomplished the collection, transmission and calculation of the Raman spectral data from CCD. The liquid colloidal gold nanoparticle (LCP-1) was used in the detection experiments of urotropine, while the characteristic peaks at 1 047 cm-1 was chosen as the identity information of urotropine. The result of detection experiments shows that the system has the ability to identify the urotropine accurately, the detection limit of urotropine standards was 001 mg·L-1, with detection limit of urotropine in yuba and rice-noodle was 05 mg·L-1. Besides, the testing time was less than 20 minutes,which meets the requirements for rapid field determination of urotropine in food.
    LI Wei, FAN Xian-guang, WANG Xin, TANG Ming, QUE Jing, HE Jian, ZUO Yong. Design of Rapid Detection System for Urotropine in Food Based on SERS[J]. Spectroscopy and Spectral Analysis, 2017, 37(6): 1778
    Download Citation