    Biao Wang, Hongfei Lu, Aoqi Li, Yue Chen, Tongxin Dai, Shuo Huang, Houquan Lian. Research of TDLAS methane detection system using VCSEL laser as the light source[J]. Infrared and Laser Engineering, 2020, 49(4): 0405002 Copy Citation Text show less
    Absorption line of CH4 molecule at 1 492.5 - 1 694.5 nm
    Fig. 1. Absorption line of CH4 molecule at 1 492.5 - 1 694.5 nm
    Experimental configuration
    Fig. 2. Experimental configuration
    2f signals’ amplitude of 10 different concentrations of CH4
    Fig. 3. 2f signals’ amplitude of 10 different concentrations of CH4
    Fitting curve of the 2f signal's amplitude and the CH4 concentration
    Fig. 4. Fitting curve of the 2f signal's amplitude and the CH4 concentration
    2f signal's amplitude (SA) at a CH4 concentration of 250 ppmv
    Fig. 5. 2f signal's amplitude (SA) at a CH4 concentration of 250 ppmv
    Noise fluctuation signal (NA) when CH4 concentration is 0 ppmv
    Fig. 6. Noise fluctuation signal (NA) when CH4 concentration is 0 ppmv
    10 hours of stability observation experimental data
    Fig. 7. 10 hours of stability observation experimental data
    Allan deviation analysis of the detection system
    Fig. 8. Allan deviation analysis of the detection system
    No.CCH4/ppmv
    50100150200250300350400450500
    152102150201251300351398454496
    251103151201253301351400452501
    34898152202249300354401450503
    45099148204255304352401451504
    54997149199254300351400451505
    647100144197251301351399453500
    750101154199251299352398451499
    853101152201251297349398452504
    951102151201252301348395448505
    105598151200247301351401449504
    Maximum absolute error5 ppmv3 ppmv6 ppmv4 ppmv5 ppmv4 ppmv4 ppmv5 ppmv4 ppmv5 ppmv
    Maximum relative error10.00%3.00%4.00%2.00%2.00%1.33%1.14%1.25%0.89%1.00%
    Table 1. Test results with N2 as background