• Acta Optica Sinica
  • Vol. 37, Issue 10, 1012001 (2017)
Shangkuo Liu1、2、*, Xun Xue1, Kun Li1, Kun Cao1, Jianke Zhao1, Yan Zhou1, and Baoli Yao1
Author Affiliations
  • 1 Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an, Shaanxi 710119, China
  • 2 University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    DOI: 10.3788/AOS201737.1012001 Cite this Article Set citation alerts
    Shangkuo Liu, Xun Xue, Kun Li, Kun Cao, Jianke Zhao, Yan Zhou, Baoli Yao. Design and Calibration of Weak-Light Single Star Simulator[J]. Acta Optica Sinica, 2017, 37(10): 1012001 Copy Citation Text show less
    Block diagram of weak-light single star simulator
    Fig. 1. Block diagram of weak-light single star simulator
    Spectral radiance calibration of the light outlet of integrating sphere
    Fig. 2. Spectral radiance calibration of the light outlet of integrating sphere
    Spectral radiance of the light outlet of integrating sphere
    Fig. 3. Spectral radiance of the light outlet of integrating sphere
    Magnitude simulation range of the weak-light single star simulator
    Fig. 4. Magnitude simulation range of the weak-light single star simulator
    Measuring illuminance in the light outlet of collimator by KLL-04 wide-range illuminometer
    Fig. 5. Measuring illuminance in the light outlet of collimator by KLL-04 wide-range illuminometer
    Star point image of high magnitude target imaging by detecting cameras
    Fig. 6. Star point image of high magnitude target imaging by detecting cameras
    Value of the 1st itemValue of the 2nd itemValue of the 4th itemValue of the 5th itemValue of the 6th item
    (5%)2(5%)2(0.2%)2(0.23%)2(0.15%)2
    Table 1. Values of the 1st、2nd、4th-6th item in equation (21)
    Diameter of star point board /mm0.0050.010.030.050.10.51
    Value of the 3rd item in equation (21)(8%)2(4%)2(1.34%)2(0.8%)2(0.4%)2(0.08%)2(0.04%)2
    Magnitude simulation range /Mv4.3-19.52.8-17.80.2-15.2-1-14-2.2-12.8-5.7-9.3-7.2-7.8
    UM /%11.608.837.827.747.707.697.69
    Table 2. Values of the 3rd item in equation (21)、simulated range and precision of magnitude
    Ample of photoelectric detector /(10-5 A)Simulated illuminance /(10-7 lx)Measured illuminance /(10-7 lx)Simulated magnitude /MvMeasured magnitude /MvAbsolute error /MvRelative error /%
    349049305400-5.67-5.77-0.101.76
    347048905350-5.66-5.76-0.101.77
    339047805230-5.64-5.74-0.101.77
    324045805010-5.59-5.69-0.101.79
    302042604690-5.52-5.62-0.101.81
    270038104200-5.40-5.50-0.101.85
    230032503570-5.22-5.32-0.101.92
    183025802830-4.97-5.07-0.102.01
    133018802070-4.63-4.73-0.102.16
    88912501390-4.19-4.30-0.112.63
    542764845-3.65-3.76-0.113.01
    301424470-3.01-3.12-0.113.65
    152215238-2.27-2.38-0.114.85
    68.696.8106-1.41-1.51-0.107.09
    26.537.441.1-0.37-0.48-0.11-
    8.1211.412.10.910.85-0.066.59
    3.374.764.621.861.900.042.15
    Table 3. Result of laboratory calibration
    Camera nameLaboratory result /MvAstro-observation result /MvAbsolute error /MvRelative error /%
    Camera 16.56.380.121.9
    Camera 215.214.80.42.6
    Table 4. Result of astro-observation