    DOI: 10.3788/IRLA20200433 Cite this Article
    Zhikun Wu, Entao Shi, Yongmei Wang. Design of PGP imaging spectrometer with eliminating spectral line bending[J]. Infrared and Laser Engineering, 2021, 50(6): 20200433 Copy Citation Text show less
    The geometry of incident ray
    Fig. 1. The geometry of incident ray
    Principle of the PGP component
    Fig. 2. Principle of the PGP component
    Smile distortion with x-field
    Fig. 3. Smile distortion with x-field
    Optical layout of improved PGP imaging spectrometer system
    Fig. 4. Optical layout of improved PGP imaging spectrometer system
    RMS spot with wavelength
    Fig. 5. RMS spot with wavelength
    Diffractive MTF curves of central wavelength
    Fig. 6. Diffractive MTF curves of central wavelength
    Smile distortion with normalized x-field
    Fig. 7. Smile distortion with normalized x-field
    Keystone distortion with wavelength
    Fig. 8. Keystone distortion with wavelength
    ParameterValue
    Spectral range/nm400-1000
    Spectral resolution/nm2.8
    Object spaceNA0.18
    Slit length/mm≥9
    Image width (spectral)/mm≤6.2
    Grating frequency/lp·mm−1360
    Table 1. Design parameters of PGP imaging spectrometer
    ParameterValue
    Bragg wavelength/nm650
    Central wavelength/nm700
    Vertex angle of the first prism/(°)1.91
    Vertex angle of the second prism/(°)14.95
    Prism materialH-K9L
    Grating materialB270
    Collimating lens focal length/mm26
    Off-axis slit position/mm2.6
    Table 2. Parameters of the improved PGP component
    ParameterValue
    Index0.0003
    Abbe0.5%
    Radius/fr2
    Thickness/mm0.02
    Decenter (surface and element)/mm0.01
    Tilt (surface and element)/(')0.5
    Irregularity0.25
    Table 3. Allowable tolerance of optical system