Zi-yue NI1、*, Da-wei CHENG2、2;, Ming-bo LIU2、2;, Yuan-bo YUE2、2;, Xue-qiang HU2、2;, Yu CHEN2、2;, and Xiao-jia LI1、1; 2; *;
Author Affiliations
  • 11. Central Iron & Steel Research Institute, Beijing 100081, China
  • 22. NCS Testing Technology Co., Ltd., Beijing 100094, China
  • show less
    DOI: 10.3964/j.issn.1000-0593(2022)04-1117-05 Cite this Article
    Zi-yue NI, Da-wei CHENG, Ming-bo LIU, Yuan-bo YUE, Xue-qiang HU, Yu CHEN, Xiao-jia LI. The Detection of Mercury in Solutions After Thermal Desorption-Enrichment by Energy Dispersive X-Ray Fluorescence[J]. Spectroscopy and Spectral Analysis, 2022, 42(4): 1117 Copy Citation Text show less
    The comparison of the spectrum for testing directly and after enrichment
    Fig. 1. The comparison of the spectrum for testing directly and after enrichment
    The test results after thermal desorption at different temperatures under three conditionsa: No mercury stabilizer, no dolomite; b: No mercury stabilizer, with dolomite; c: With mercury stabilizer, with dolomite
    Fig. 2. The test results after thermal desorption at different temperatures under three conditions
    a: No mercury stabilizer, no dolomite; b: No mercury stabilizer, with dolomite; c: With mercury stabilizer, with dolomite
    The optimization of the measure time(a): Thermal desorption; (b): X-ray fluorescence analyzer
    Fig. 3. The optimization of the measure time
    (a): Thermal desorption; (b): X-ray fluorescence analyzer
    Relationship between magnification and injection volume
    Fig. 4. Relationship between magnification and injection volume
    The influence of injection volume
    Fig. 5. The influence of injection volume
    The influence of pumping rate
    Fig. 6. The influence of pumping rate
    The calibration curve of mercury
    Fig. 7. The calibration curve of mercury
    测试次数测试结果/(μg·mL-1)
    10.046
    20.045
    30.047
    40.047
    50.048
    60.048
    70.047
    80.045
    90.050
    100.044
    110.049
    SD/(μg·mL-1)0.002
    RSD/%4.048
    Table 1. Stability experiments
    测试结果/
    (μg·mL-1)
    干扰离子浓度/(μg·mL-1)
    00.55
    10.0520.0500.051
    20.0500.0530.047
    30.0540.0480.047
    Table 2. The influence of interfering ions
    测试方法1#2#3#4#
    ICP-MS/(ng·mL-1)0.5570.061ND0.068
    加标前/(μg·mL-1)NDNDNDND
    XRF加标后/(μg·mL-1)0.0500.0490.0470.051
    加标回收率/%100.798.994.3102.6
    Table 3. The recovery rate