• Spectroscopy and Spectral Analysis
  • Vol. 42, Issue 2, 622 (2022)
Yi-chuan TANG1、*, Yan-jie CUI1、1;, Jian-ying ZHANG1、1;, Sheng HE2、2;, Tao ZHOU1、1; *;, and Bing WU1、1;
Author Affiliations
  • 11. National Institute of Metrology, China, Beijing 100029, China
  • 22. Beijing Research Institute of Uranium Geology, Beijing 100029, China
  • show less
    DOI: 10.3964/j.issn.1000-0593(2022)02-0622-05 Cite this Article
    Yi-chuan TANG, Yan-jie CUI, Jian-ying ZHANG, Sheng HE, Tao ZHOU, Bing WU. Determination of Trace Oxygen in High Purity Gold by Using Inert Gas Fusion Infrared Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 622 Copy Citation Text show less
    Results of Oxygen in gold(a): 1st measurement of gold; (b): 2nd measurement of gold;(c): measurement of gold and tin
    Fig. 1. Results of Oxygen in gold
    (a): 1st measurement of gold; (b): 2nd measurement of gold;(c): measurement of gold and tin
    Trend of results of oxygen in gold with different mass of tin particle
    Fig. 2. Trend of results of oxygen in gold with different mass of tin particle
    Linear calibration curve of oxygen
    Fig. 3. Linear calibration curve of oxygen
    No.ResultsNo.Results
    10.06870.079
    20.07180.060
    30.08090.079
    40.050100.067
    50.080110.066
    60.083120.067
    平均值0.07
    标准偏差0.01
    定量限0.1
    Table 1. Results of oxygen blank in tin particle (mg·kg-1, n=12)
    No.氧含量测量值
    /(mg·kg-1)
    No.氧含量测量值
    /(mg·kg-1)
    10.98470.957
    21.09681.117
    31.10991.224
    40.905101.103
    51.170111.058
    60.987121.327
    AVE1.09
    RSD/%12
    校正值1.1
    Table 2. Measurement results of oxygen in high purity gold
    No.金质量/g加标量/μg测得总量/μg回收率/%
    10.49814.85417.681103
    20.5044.3026.66899
    30.50547.59652.635105
    40.5129.96312.30499
    50.50137.08939.737101
    Table 3. Measurement results of oxygen in high purity gold by using addition method
    No.SRM685w
    /cps
    High purity
    gold/cps
    High purity gold
    /(mg·kg-1)
    1279 074152 4831.10
    2321 085136 7960.98
    3183 148144 0831.04
    4338 216133 0160.96
    5356 608114 4020.82
    6234 436113 0900.81
    7298 16483 7180.60
    8241 209150 7881.08
    9268 043132 5030.95
    10250 66094 8140.68
    11286 967111 3750.80
    12281 598152 4831.10
    AVE278 267126 6290.91
    RSD/%171818
    Table 4. Measurement results of oxygen in high purity gold by using SIMS method
    Yi-chuan TANG, Yan-jie CUI, Jian-ying ZHANG, Sheng HE, Tao ZHOU, Bing WU. Determination of Trace Oxygen in High Purity Gold by Using Inert Gas Fusion Infrared Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 622